![Thermo Scientific SOLA II Flare Скачать руководство пользователя страница 118](http://html1.mh-extra.com/html/thermo-scientific/sola-ii-flare/sola-ii-flare_user-manual_1105273118.webp)
Appendix A
Watlow EZ Zone PM6 Series
Table A–23. Oven settings: Setup Page > Global 1
Parameter
Default
Comment
Display
Software
Display
Software
Display Units
C
Display units are in °C.
AC Line Frequency
60 Hz
The frequency to the applied AC line
power source.
Communications LED Action
Both
Both comm port 1 and 2 selected.
Zone
Off
Channel
On
The Channel LED is on.
Display Pairs
1
The number of display pairs.
Display Time
0
The time delay for toggling between
channel 1 and channel 2.
User Settings Save
None
User Settings Restore
None
Table A–24. Oven settings: Operations Page > Analog Input 1
Parameter
Description
Display
Software
Analog Input Value
The process value (°C).
Input Error
The cause of the most recent error. Possible displays listed below.
None
Open
Fail
Shorted
Measurement Error
Bad Calibration Data
Ambient Error
RTD Error
Not Sourced
Calibration Offset
The offset applied to the input reading to compensate factors that cause the input reading to
vary from the actual process value (°C).
Default = 0.
Thermo Scientific
SOLA II Flare User
Manual
A-14
Содержание SOLA II Flare
Страница 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Страница 2: ......
Страница 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 42: ...Chapter 3 Start Up Shutdown Thermo Scientific SOLA II Flare User Manual 3 6 Figure 3 4 Diagnostics Menu...
Страница 44: ......
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 80: ......
Страница 104: ......
Страница 122: ......
Страница 136: ......
Страница 148: ......
Страница 158: ......
Страница 160: ......
Страница 162: ......