Chapter 5
Configuration
4. Select Disabled or one of 10 predefined control output states.
0. Disabled
1. Local Low Cal A
2. Local Low Cal B
3. Local Mid Cal A
4. Local Mid Cal B
5. Local High Cal B
6. Probe Low Cal A
7. Probe Low Cal B
8. Probe Mid Cal A
9. Probe Mid Cal B
10. Probe High Cal B
V a l i d a t i o n
S e t u p
S t e p
1
S e l e c t i o n
∨
∧
E n t
E x i t
Figure 5-50.
5. Enter the number of minutes (0-60) that the SOLA II Flare is to remain
in this particular validation step.
V a l i d a t i o n
S e t u p
S t e p
1
T i m e
∨
∧
E n t
E x i t
Figure 5-51.
Thermo Scientific
SOLA II Flare User
Manual
5-19