Chapter 1
Product Overview
Thermo Scientific
SOLA II Flare User Manual
1-7
Mixing Chamber
The mixing chamber mixes the sample and air before they enter the pyrolyzer.
Figure 1-3. Typical mixing chamber
Dual Measuring Range
All SOLA II Flare systems utilize a dual measuring range configuration. The
dual ranges are achieved by using two sample injection valves to achieve a 100:1
dilution ratio. The two injection valves allow for two completely separate
measuring ranges in the PUVF Detector.
The Low Range uses a 10-port sample injection valve with a 100μ
sample loop and the low measurement range calibrated in the detector.
The High Range uses a 6-port sample injection valve with a 1μ sample
loop and the high measurement range calibrated in the detector.
The results of the sulfur concentration level being measured by the SOLA II
Flare are selected by the auto-ranging function.
Figure 1-4. Dual inject valves (Valco) w/o calibration valve in oven
Содержание SOLA II Flare
Страница 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Страница 2: ......
Страница 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 42: ...Chapter 3 Start Up Shutdown Thermo Scientific SOLA II Flare User Manual 3 6 Figure 3 4 Diagnostics Menu...
Страница 44: ......
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 80: ......
Страница 104: ......
Страница 122: ......
Страница 136: ......
Страница 148: ......
Страница 158: ......
Страница 160: ......
Страница 162: ......