Appendix C: TCP/IP Bridge
The SOLA II Flare Modbus TCP/IP option is a microprocessor-driven board
that provides a bridge between the SOLA II Flare standard serial Modbus and a
TCP/IP Modbus network using 10/100 Base-T. This function is provided by
the NetBurner™.
Hardware
Configuration
•
TCP/IP Bridge board installed inside the SOLA II Flare enclosure
•
SOLA II Flare Modbus configuration must be set to 38400 baud with
Modbus ID 1
•
Bridge board has a standard RJ45 connector for the 10/100 Base-T
cable
Figure C–1. TCP/IP Ethernet bridge board, P/N 55-1228-0 (newer models)
Software
Configuration
As shipped, the Bridge defaults to an IP address of 00.00.00.00. At power- up
(with an IP address of 00.00.00.00), the Bridge attempts to negotiate an IP
through a DHCP server. If a DHCP server is not available or if it is desired to
assign the instrument a static IP, use the provided utility program to set up the
IP address. Perform the following steps to search for the Bridge on the network
and assign an IP address.
1. Locate the IPSetup.exe utility that is included with the disk that ships
with the TCP/IP Bridge option.
Thermo Scientific
SOLA II Flare User
Manual
C-1
Содержание SOLA II Flare
Страница 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Страница 2: ......
Страница 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 42: ...Chapter 3 Start Up Shutdown Thermo Scientific SOLA II Flare User Manual 3 6 Figure 3 4 Diagnostics Menu...
Страница 44: ......
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 80: ......
Страница 104: ......
Страница 122: ......
Страница 136: ......
Страница 148: ......
Страница 158: ......
Страница 160: ......
Страница 162: ......