![Thermo Scientific SOLA II Flare Скачать руководство пользователя страница 129](http://html1.mh-extra.com/html/thermo-scientific/sola-ii-flare/sola-ii-flare_user-manual_1105273129.webp)
Appendix B
Rotary Valve Service
Conditioning
Procedure for High
Temperature Valves
High temperature valves require conditioning when the rotor is replaced. If a
high temperature valve (WT or UWT series) is used at less than 300°C, it may
become sticky or difficult to turn. This tendency can usually be eliminated by
repeating the conditioning procedure that is initially done at the factory. With
carrier gas (oxygen-free) flowing through all the ports, rapidly heat the valve to
325°C. When this temperature is reached, cycle the valve 10 times and let it
cool to operating temperature.
Two-Position Air
Actuator O-Ring
Replacement
You will need the following to perform the tasks described in this section.
•
9/64” hex driver
•
3/8” open-end wrench
•
3/16” screwdriver
•
An awl or small jeweler’s screwdriver
•
Silicone lubricant (such as Dow Corning® DC-111)
•
Lint-free tissues and a clean shop rag
•
Standard O-Ring kit (VICI P/N OR)
•
High temperature O-Ring kit (VICI P/N ORT)
Disassembly
1. Apply air pressure to the actuator inlet nearest the valve. Then use the
open-end wrench to remove the air supply lines from the actuator.
Figure B–5.
2. Remove the valve and valve-mounting hardware from the actuator (as
shown in Figure B–5):
a. Use the 9/64" hex driver to loosen the HWSC-SC8-10B/socket-
head screw in the black anodized CR3/clamp ring on the actuator.
b. Pull off the standoff with the valve attached.
c. Use the 9/64” hex driver to remove the two HWSC-SC8-6/socket-
head screws that hold the clamp ring to the actuator.
Thermo Scientific
SOLA II Flare User
Manual
B-7
Содержание SOLA II Flare
Страница 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Страница 2: ......
Страница 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 14: ......
Страница 18: ......
Страница 42: ...Chapter 3 Start Up Shutdown Thermo Scientific SOLA II Flare User Manual 3 6 Figure 3 4 Diagnostics Menu...
Страница 44: ......
Страница 48: ......
Страница 70: ......
Страница 72: ......
Страница 80: ......
Страница 104: ......
Страница 122: ......
Страница 136: ......
Страница 148: ......
Страница 158: ......
Страница 160: ......
Страница 162: ......