12
Measure Protein Lowry
118
NanoDrop One User Guide
Thermo Scientific
Settings for Protein Lowry Measurements
To show the Protein Lowry settings, from the Protein Lowry measurement screen, tap
>
Protein Lowry Setup.
Related Topics
•
Note
You can edit the Curve Type setting when measuring standards by changing the list
box at the top of the application measurement screen. You can edit the concentration
value for a standard from the application setup screen. After the first sample measurement,
these settings cannot be changed.
Setting
Description
Curve Type
Specify type of equation used to create standard curve from standard concentration values.
Available options:
–
Linear
: Draws the linear least squares line through all measured standards (requires
reference measurement and at least one standard)
–
Interpolation
: Draws a series of straight lines to connect all measured standards
(requires reference measurement and at least one standard)
–
2
nd
order polynomial
: Draws the 2
nd
order least squares polynomial using all
measured standards (requires reference measurement and at least two standards)
–
3
rd
order polynomial
: Draws the 3
rd
order least squares polynomial using all
measured standards (requires reference measurement and at least three standards)
Replicates
Enter number of measurements of the reference or the same standard or sample that are
averaged together to produce its associated concentration value.
Note
: Replicates setting cannot be changed after the first standard has been measured.
Standards
Enter actual concentration value of each standard.
Note
: Concentration values can be entered in any order but the standards must be measured
in the order they were entered.
Содержание NanoDrop One
Страница 1: ...NanoDrop Micro UV Vis Spectrophotometers NanoDrop One User Guide 269 309102 Revision A May 2017 ...
Страница 18: ......
Страница 28: ......
Страница 58: ......
Страница 134: ...13 Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Страница 144: ......
Страница 152: ......
Страница 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Страница 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Страница 193: ...18 Learning Center Set Up the Instrument Thermo Scientific NanoDrop One User Guide 187 Instrument Settings ...
Страница 293: ...18 Learning Center Multimedia Thermo Scientific NanoDrop One User Guide 287 Pedestal Cleaning and Reconditioning ...
Страница 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Страница 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Страница 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...