9
Measure Protein A205
86
NanoDrop One User Guide
Thermo Scientific
Related Topics
Sample Pathlength
• For micro-volume measurements, the software selects the optimal
pathlength (between 1.0 mm and 0.03 mm) based on sample
absorbance at the analysis wavelength.
• For cuvette measurements, pathlength is determined by the cuvette
Pathlength setting in the software (see
• Displayed spectra and absorbance values are normalized to a 10 mm
pathlength equivalent.
Reported Values
•
Protein concentration.
Reported in selected unit (mg/mL or μg/mL).
Calculations are based on Beer-Lambert equation using corrected
protein absorbance value.
Содержание NanoDrop One
Страница 1: ...NanoDrop Micro UV Vis Spectrophotometers NanoDrop One User Guide 269 309102 Revision A May 2017 ...
Страница 18: ......
Страница 28: ......
Страница 58: ......
Страница 134: ...13 Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Страница 144: ......
Страница 152: ......
Страница 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Страница 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Страница 193: ...18 Learning Center Set Up the Instrument Thermo Scientific NanoDrop One User Guide 187 Instrument Settings ...
Страница 293: ...18 Learning Center Multimedia Thermo Scientific NanoDrop One User Guide 287 Pedestal Cleaning and Reconditioning ...
Страница 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Страница 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Страница 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...