10
Measure Protein BCA
Thermo Scientific
NanoDrop One User Guide
93
• After the minimum number of standards has been measured for the selected curve type,
the message “Invalid Curve” changes to “Valid Curve.” (This occurs even when additional
standards have been defined but not yet measured.) If the “Invalid Curve” message
remains after all entered standards have been measured, try:
–
selecting a different curve type
–
remeasuring standards using the correct standard material
Valid Curve indicator
: This is only an indicator that the required minimum number of
points has been established for the selected curve type. It does not validate the integrity of
the curve. For example, additional standards may be required to cover the expected assay
concentration range.
To measure Protein BCA standards and samples
Before you begin...
Before taking pedestal measurements with the NanoDrop One instrument, lift the instrument
arm and clean the upper and lower pedestals. At a minimum, wipe the pedestals with a new
laboratory wipe. For more information, see
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into
the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently
damage the quartz fiber optic cables.
To measure Protein BCA standards and samples
1. From the Home screen, select the
Proteins
tab and tap
Protein BCA
.
2. Specify a
and enter the
concentration of each standard
Tip
: For this assay, we recommend setting
Curve Type
to “Linear”.
3. Measure blank:
–
pipette 2 μL DI H
2
O onto lower pedestal and lower arm, or insert DI H
2
O
blanking cuvette into cuvette holder
Содержание NanoDrop One
Страница 1: ...NanoDrop Micro UV Vis Spectrophotometers NanoDrop One User Guide 269 309102 Revision A May 2017 ...
Страница 18: ......
Страница 28: ......
Страница 58: ......
Страница 134: ...13 Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Страница 144: ......
Страница 152: ......
Страница 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Страница 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Страница 193: ...18 Learning Center Set Up the Instrument Thermo Scientific NanoDrop One User Guide 187 Instrument Settings ...
Страница 293: ...18 Learning Center Multimedia Thermo Scientific NanoDrop One User Guide 287 Pedestal Cleaning and Reconditioning ...
Страница 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Страница 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Страница 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...