Before you begin
First make sure the pedestals are properly conditioned. To test pedestal conditioning, clean the
pedestals with a new laboratory wipe, then pipette 1 μL DI H
2
O onto the lower pedestal. The
droplet should “bead up” as shown below. If it does not,
.
To run performance verification
1. From the instrument home screen, tap
(Diagnostics) and then tap
Performance Verification
.
A message asks for target absorbance values.
2. Enter each lot-specific target absorbance value from the label on the PV-1 ampoule in its
associated entry box and then tap
Done
.
3. Lift the instrument arm and clean the upper and lower pedestal with a new laboratory
wipe.
Droplet “beads up” on
properly conditioned pedestal
Tap an entry box to display
a numerical keyboard
Содержание NanoDrop One
Страница 1: ...NanoDrop Micro UV Vis Spectrophotometers NanoDrop One User Guide 269 309102 Revision A May 2017 ...
Страница 18: ......
Страница 28: ......
Страница 58: ......
Страница 134: ...13 Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Страница 144: ......
Страница 152: ......
Страница 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Страница 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Страница 193: ...18 Learning Center Set Up the Instrument Thermo Scientific NanoDrop One User Guide 187 Instrument Settings ...
Страница 293: ...18 Learning Center Multimedia Thermo Scientific NanoDrop One User Guide 287 Pedestal Cleaning and Reconditioning ...
Страница 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Страница 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Страница 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...