Safety Information
Thermo Scientific
NanoDrop One User Guide
309
Most solvents typically used in life science laboratories are compatible with the fiber optic
pedestals of all NanoDrop spectrophotometers. However, the high vapor pressure properties
of some solvents may not be conducive to small volume measurements when using the
pedestal for measurements on any of the NanoDrop instruments. If you are measuring
samples with high vapor pressures, use an instrument with provision for measuring samples in
cuvettes.
The following solvents are compatible for use on the
pedestals
of all NanoDrop instruments.
It is recommended that all corrosive solvents be wiped from the pedestal immediately upon
completion of a measurement. It is also recommended that the user end a series of
measurements with a dH
2
O sample to ensure that solvents are not inadvertently left on the
pedestal.
The diaphragm around the pedestal of the NanoDrop is permanently affixed to the
instrument. Do not attempt to remove the diaphragm or break the seal. Avoid prolonged
exposure of the diaphragm to HCl, alcohol, bleach, acetone or other solvents as the adhesive
securing the seal may be affected. If the seal comes loose please contact us.
Biohazard or Radioactive Materials and Infectious Agents
Biological samples such as tissues, body fluids, infectious agents, and blood of humans and
other animals have the potential to transmit infectious diseases. Wear appropriate protective
equipment. Individuals should be trained according to applicable regulatory and organization
requirements before working with potentially infectious materials. Follow your organization’s
Biosafety Program protocols for working with and/or handling potentially infectious
materials.
NOTICE
Spillage of these solvents on surfaces other than the pedestals may damage the
instrument.
• methanol
• ethanol
• n-propanol
• isopropanol
• butanol
• acetone
• ether
• chloroform
• carbon tetrachloride
• DMSO
• DMF
• acetonitrile
• THF
• toluene
• hexane
• benzene
• sodium hydroxide
• sodium hypochlorite (bleach)
• dilute HCl
• dilute HNO
3
• dilute acetic acid
NOTICE
All forms of Hydrofluoric Acid (HF) are incompatible as the fluoride ion will
etch the fiber optic cable.
Содержание NanoDrop One
Страница 1: ...NanoDrop Micro UV Vis Spectrophotometers NanoDrop One User Guide 269 309102 Revision A May 2017 ...
Страница 18: ......
Страница 28: ......
Страница 58: ......
Страница 134: ...13 Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Страница 144: ......
Страница 152: ......
Страница 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Страница 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Страница 193: ...18 Learning Center Set Up the Instrument Thermo Scientific NanoDrop One User Guide 187 Instrument Settings ...
Страница 293: ...18 Learning Center Multimedia Thermo Scientific NanoDrop One User Guide 287 Pedestal Cleaning and Reconditioning ...
Страница 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Страница 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Страница 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...