11
Measure Protein Bradford
104
NanoDrop One User Guide
Thermo Scientific
Related Topics
• Working with standard curves
• Best practices for protein measurements
• Measure a Micro-Volume Sample
Tip
: If using a cuvette, make sure to
with instrument light
path.
–
tap
Blank
and wait for measurement to complete
–
lift arm and clean both pedestals with new laboratory wipe, or remove cuvette
4. Measure reference standard:
–
pipette 2 μL reference solution onto pedestal, or insert reference cuvette (reference
solution should contain none of the standard protein stock, see
for details)
–
lower arm to start measurement (or tap
Measure
if Auto-Measure is off )
–
lift arm and clean both pedestals with new wipe, or remove cuvette
–
if Replicates setting is greater than 1, repeat measurement
5. Measure remaining standards:
–
pipette 2 μL standard 1 onto pedestal, or insert standard 1 cuvette
–
lower arm to start measurement (or tap
Measure
if Auto-Measure is off )
–
lift arm and clean both pedestals with new wipe, or remove cuvette
–
if Replicates setting is greater than 1, repeat measurement
–
repeat substeps above for each additional standard (when specified number of
standards and replicates have been measured, a message asks whether to load more
standards or begin measuring samples)
–
if finished measuring standards, tap
Done
(swipe left to view standard curve)
6. Measure samples:
–
pipette 2 μL sample 1 onto pedestal, or insert sample 1 cuvette
–
lower arm to start measurement (or tap
Measure
if Auto-Measure is off )
–
lift arm and clean both pedestals with new wipe, or remove cuvette
–
if Replicates setting is greater than 1, repeat measurement
7. When you are finished measuring samples, tap
End Experiment
.
8. Lift the arm and clean both pedestals with a new wipe, or remove the sample cuvette.
Содержание NanoDrop One
Страница 1: ...NanoDrop Micro UV Vis Spectrophotometers NanoDrop One User Guide 269 309102 Revision A May 2017 ...
Страница 18: ......
Страница 28: ......
Страница 58: ......
Страница 134: ...13 Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Страница 144: ......
Страница 152: ......
Страница 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Страница 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Страница 193: ...18 Learning Center Set Up the Instrument Thermo Scientific NanoDrop One User Guide 187 Instrument Settings ...
Страница 293: ...18 Learning Center Multimedia Thermo Scientific NanoDrop One User Guide 287 Pedestal Cleaning and Reconditioning ...
Страница 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Страница 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Страница 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...