Instrument Diagnostics
306
NanoDrop One User Guide
Thermo Scientific
After the tenth measurement, a message indicates whether the instrument passed or failed
performance verification:
8. If the instrument failed, immediately repeat step 6 using ten 2 μL aliquots of the PV-1
solution.
9. When finished, tap
End Experiment
and
clean the pedestals with 3–5 μL DI H
O
.
After the test is completed, the results are available from the Data Viewer (see example
below). See
Manage identifiers on the instrument
for details.
To interpret performance verification results
If your instrument failed performance verification and you repeated ten measurements using
2 uL aliquots,
Related Topics
Содержание NanoDrop One
Страница 1: ...NanoDrop Micro UV Vis Spectrophotometers NanoDrop One User Guide 269 309102 Revision A May 2017 ...
Страница 18: ......
Страница 28: ......
Страница 58: ......
Страница 134: ...13 Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Страница 144: ......
Страница 152: ......
Страница 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Страница 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Страница 193: ...18 Learning Center Set Up the Instrument Thermo Scientific NanoDrop One User Guide 187 Instrument Settings ...
Страница 293: ...18 Learning Center Multimedia Thermo Scientific NanoDrop One User Guide 287 Pedestal Cleaning and Reconditioning ...
Страница 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Страница 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Страница 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...