64
3. Select the empty filter position by pressing the
Right
arrow key until
the filter wheel on the display rotates to the empty position, and the
text “
Empty
” is shown.
to
4. Press the
OK
button. Enter the wavelength of the filter, for example,
492
, with the number keys, and then press the
OK
button.
5. Press the
F1
key to accept the filter settings.
Note!
The instrument starts the filter initialization.
6. Press the
Right
arrow key to return to the
Main
menu.
Removing a filter
You can remove a filter from the instrument settings by selecting the position and pressing the
F3
(Remove)
key.
Note!
The filter must physically be removed from the filter wheel.
Printer
You can add a header in the
Printer
parameter window, for example, the laboratory information, to all
of the reports to be printed. You can do this by selecting
Yes
for the
Enable header
.
The maximum length of the header is about 50 characters.
Startup
In the
Startup
parameters, the
Plate position
setting defines the plate carrier position after the
instrument is switched on.
You can set the startup temperature. The instrument starts to warm up to the set temperature
immediately after it has been switched on.