Note!
The main parameters are locked when a run (measured data) for this protocol exists.
Chapter 7: Protocol Creation Examples
This section provides examples on how to open a new protocol, how to program main parameters
(wavelength, shaking, and incubation), the plate layout and precalculation, calculation and
interpretation parameters, and how to save the protocol.
Note!
The
Incubation
row is only active with the Multiskan FC with incubator model.
Note!
Ensure that you save the changes you make in each step. Refer to “Saving a new
(active) protocol”.
Opening a new protocol
1. Press the
FILE
key in the
Main
menu.
2. Select
New
and press the
OK
button.
Programming a wavelength
1. Select the
Measurement
row in the
Main
menu using the
Down
arrow key
and press the
OK
button.
2. Press the
OK
button at the
Filter 1 (nm)
item and select, for example,
450
as the filter 1 value using the
Down
arrow key, and then press the
OK
button.
3. Press the
F1
key to accept the selection and return to the
Main
menu.
Programming shaking parameters
1. Select the
Shaking
row in the
Main
menu using the
Down
arrow key and
press the
OK
button.
2. Press the
OK
button at the
Mode
item and select, for example,
Before
measurement
as the shaking mode using the
Down
arrow key, and then
press the
OK
button.