Ion S5
™
Controls Kit Plus
The Ion S5
™
Controls Kit Plus (Cat. No.
) is ordered separately. See the
Ion S5
™
Controls Kit Plus
Product Information Sheet
(Pub. No. MAN0016206) for details.
Contents
Amount
Storage
Human CEPH Genomic DNA Control (red cap)
30 µL
−30°C to −10°C
Human CEPH Control 200 Library (yellow cap)
12 µL
E. coli
DH10B Control 400 Library (orange cap)
16 µL
E. coli
DH10B Control 600 Library
The
E. coli
DH10B Control 600 Library includes sufficient material for 4 templating runs on the
Ion Chef
™
System. The
E. coli
DH10B Control 600 Library (Cat. No. A32470) is ordered separately.
Contents
Amount
Storage
E. coli
DH10B Control 600 Library (lilac cap)
20 µL
–30°C to –10°C
Required materials not supplied
Unless otherwise indicated, all materials are available through
. "MLS" indicates that
the material is available from
or another major laboratory supplier.
Item
Source
Ion Chef
™
System
Microcentrifuge
[1]
MLS
2
‑
, 20
‑
, 200
‑
, and 1,000
‑
µL pipettes and appropriate filtered tips
MLS
Microcentrifuge tubes, 1.5-mL or 1.7-mL
MLS
Vortex mixer with a rubber platform
MLS
Gloves, powder-free nitrile
MLS
Ice buckets and ice
—
Nuclease-free water, molecular biology grade
MLS
Isopropyl alcohol, 70% solution
MLS
Wipes, disposable lint-free
MLS
(Optional)
Uninterruptible Power Supply (UPS)
[2]
MLS
[1]
Must fit standard 0.2- and 1.5
‑
mL microcentrifuge tubes and generate 21,000 ×
g
. To convert the RPMs of your centrifuge to RCF in
units of gravity, see
tools.thermofisher.com/content/sfs/brochures/TR0040-Centrifuge-speed.pdf
[2]
For laboratories that experience frequent power outages or line voltage fluctuations, we recommend that you use an uninterruptible
power supply that is compatible with 2500 W output or higher.
Chapter 1
Product information
Required materials not supplied
1
Ion 520
™
& Ion 530
™
ExT Kit − Chef User Guide
9
Содержание Ion 520