Troubleshooting
■
System setup and operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
■
S5 Systems setup and operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Ion Chef
™
System setup and operation
Observation
Possible cause
Recommended action
Instrument does not display
the home screen when
powered on
Multiple causes are possible.
1. Power off the instrument, wait
30 seconds, then power on the
instrument.
2. If the instrument fails again, contact
Technical Support.
The instrument door cannot be
opened
An obstruction was present on
or around the door mechanism.
Remove the obstruction blocking the door,
then operate the instrument normally.
A hardware or software error
occurred.
Contact Technical Support to report the
problem and for further assistance.
The instrument stops during a
run
The instrument encountered an
internal error.
1. Record the error displayed on the
instrument display, then tap
OK
.
2. Contact Technical Support to report the
problem and for further assistance.
A consumable was improperly
loaded.
Ensure that all consumables are loaded
according to the instructions provided.
Liquid residue is present in the
Recovery centrifuge following a
run
During normal instrument
operation, a noticeable coating
of liquid can collect on the
bowl and buckets of the
Recovery centrifuge following
repeated runs.
Remove the residue as instructed in “Clean
the Ion Chef
.
A