3.
Load the cartridge into the Reagents station so that it snaps into place and is level on the deck.
IMPORTANT!
Do not force the Ion Chef
™
cartridges into place. Each cartridge fits only one
location on the deck and in one orientation. If a cartridge does not fit, confirm that you are loading
the correct cartridge in the correct orientation.
A
B
C
D
E
F
G
H
1
2
3
4
5
6
7
8
9
10
11 12
1
2
1
Reagents station (4°C)
2
Ion S5
™
ExT Chef Reagents cartridge
4.
Uncap, then load the Ion Chef
™
Library Sample Tube containing 50 µL of diluted library into
Position A on the Reagents cartridge.
A
00012216
1
D
4
C
3
B
2
1
Position A (Library Sample Tube)
2
Position B (2 M NaOH)
3
Position C (Ion S5
™
ExT Pellet)
4
Position D (Ion S5
™
ExT Pellet)
IMPORTANT!
Orient the Ion Chef
™
Library Sample Tube so that the barcode is visible and
oriented to the right.
5.
Uncap the tube of 2 M NaOH in Position B and the tubes in Positions C and D, each containing an
Ion S5
™
ExT Pellet.
IMPORTANT!
When the Reagents cartridge is loaded:
·
Press down on the Library Sample Tubes to ensure that it is firmly seated in the cartridge.
·
Ensure that
all
four tubes are uncapped.
6.
Gently tap the Ion S5
™
Chef Solutions cartridge on the bench to force the reagents to the bottoms
of the tubes.
Chapter 4
Start the Ion Chef
™
run
Load the Ion Chef
™
System
4
30
Ion 520
™
& Ion 530
™
ExT Kit − Chef User Guide