5.
Slide the catch forward to allow the locking bracket to pivot upward. Load the Ion S5
™
ExT Tip
Cartridge into the
New
Pipette Tip Position, pull the bracket downward, then push the catch
backward to lock the cartridge in place.
2
3
1
4
1
New Pipette Tip Position
2
Bracket
3
Catch
4
New Ion S5
™
ExT Tip Cartridge
6.
Load a new PCR Plate into the thermal cycler sample block, then insert a new Frame Seal v2
underneath the automated heated cover.
IMPORTANT!
When the Frame Seal v2 is positioned correctly, its tabs project upward and
contact the heated cover.
A
B
C
D
E
F
G
H
1
2
3
4
5
6
7
8
9
10
11 12
A
B
C
D
E
F
G
H
1
2
3
4
5
6
7
8
9
10
11
12
4
1
3
2
1
Thermal cycler sample block
2
Well A1
3
Cover
4
Keyed corner
Load the Reagents and Solutions cartridges
IMPORTANT!
Thaw the Reagents cartridge at room temperature for 45 minutes before use.
1.
Gently tap the Ion S5
™
ExT Chef Reagents cartridge on the bench to force the reagents to the
bottoms of the tubes.
2.
If bubbles are present below the surface of the liquid, repeat step 1.
Chapter 4
Start the Ion Chef
™
run
Load the Ion Chef
™
System
4
Ion 520
™
& Ion 530
™
ExT Kit − Chef User Guide
29