About the Ion Chef
™
System touchscreen interface
The Ion Chef
™
System features a simple interface for loading chips, cleaning the instrument, and
performing system maintenance and configuration tasks.
2
1
3
4
5
1
Set up run
button – Set up the Ion Chef
™
template preparation and chip-loading routine. Choose
Step by Step
to
have the instrument lead you stepwise through installation of reagents and consumables, or choose
Quick Start
to
proceed if you have already installed the consumables.
2
Open Door
button.
3
Notifications
button – View notifications about instrument status during and between runs.
4
Quick Start
button – Proceed directly to the Quick Start instrument setup mode. User verifies the loading of a new
pipette tip cartridge and an empty pipette tip rack to hold waste tips generated during the run, before proceeding to
Deck Scan.
5
Settings
button – Advance to a screen providing the following options:
•
Notifications
: view notifications about instrument status during and between runs
•
Instrument Settings
: view current settings and network configuration, set instrument name, adjust time zone
•
Service tools
: access screens for service-related maintenance and instrument diagnostics
•
Torrent Server
: add and manage Ion Torrent
™
Server connections
•
Clean Ion Chef
: proceed directly to the instrument cleaning routine
•
Check for updates
: check availability of system software updates
Chapter 1