g.
Enter the appropriate number of flows in the
Flows
field.
Sequencing type
Number of flows
400
‑
base
‑
read
850
600
‑
base
‑
read
1,350
IMPORTANT!
Do not exceed 1,400 total flows for one Ion S5
™
ExT Sequencing Reagents
cartridge.
h.
Select or edit the remaining optional information fields appropriately for your run.
IMPORTANT!
For
de novo
sequencing applications that do not include a reference BAM
file, select
Enable Calibration Standard
from the
Base Calibration Mode
dropdown list in
the
Advanced Settings
pane. Select this option only if you will add Ion S5
™
ExT Calibration
Standard to the Ion Chef
™
Library Sample Tube. Selecting this option and adding the Ion S5
™
ExT Calibration Standard to your library allows greater accuracy of base-calling in libraries for
which a reference BAM file does not exist.
i.
Click
Next
.
5.
Review the
Plugins
and
Projects
screens, then make selections appropriate to your run.
6.
In the
Plan
step, enter or make the following selections:
a.
Enter a Run Plan Name, then select Reference and BED files appropriate to your run.
Chapter 3
Create a Planned Run
Create a Planned Run
3
Ion 520
™
& Ion 530
™
ExT Kit − Chef User Guide
19