1.
After completion of initialization, tap
Run
in the instrument touchscreen. The door and chip clamp
unlock.
2.
Remove the used sequencing chip, then fasten a chip that is loaded with template-positive Ion
Sphere
™
Particles in the chip clamp.
3.
Push the chip clamp all the way in to engage, close the instrument door, then tap
Next
.
Note:
Do not remove the chip from the chip clamp until completion of the run. Removing and
reinserting the chip risks introducing air bubbles in the chip.
4.
Confirm that the correct Planned Run has auto-populated, then tap
Review
.
Note:
Ensure the
Enable post-run clean
checkbox is selected so that the post-run cleaning is
performed automatically.
Chapter 6
Start the sequencing run
Start the sequencing run
6
50
Ion 520
™
& Ion 530
™
ExT Kit − Chef User Guide