Ion Chef
™
pre-run checklist
A
B
C
D
E
F
G
H
1
2
3
4
5
6
7
8
9
10
11 12
530
DABD02226
Ion
Che
f
™
S5 Series
Chip
Balance
1) A new Ion S5™ ExT Tip Cartridge
is loaded in the New Pipette Tip position.
2) Bracket is pulled downward to lock the
tip cartridge in place.
1) A new PCR Plate is loaded in the thermal
cycler sample block.
2) A Frame Seal v2 is in place and oriented
under the heated cover.
1) The tip rack from previous run is transferred
from the New Pipette Tip position to the
Used Pipette Tip position.
1) Recovery Tubes (v2) are securely seated
in the Recovery centrifuge buckets.
2) Recovery centrifuges are load-balanced.
3) Recovery Station Disposable Lids (v2) are
positioned so that the port is oriented
toward the rear of the instrument.
1) Enrichment Cartridge v2 is pressed into place
at the Enrichment station so that the cartridge
is
firmly seated
and
level
with the deck.
2) Lettering on the cartridge is right-side up and
positioned to the right of the enrichment tubes.
1) Adapter/chip assembly is correctly
seated in the Chip-loading centrifuge bucket
loaded in Position 1, with the adaptor clips
inserted into the bucket slots.
2) Centrifuge buckets containing the adapter/chip
assembly and Chip Balance, loaded in Position 2,
are securely seated in the centrifuge rotor, and
freely pivot 90º outwards.
1) Solutions cartridge is pressed into
place at the Solutions station so that the
cartridge is
firmly seated
and
level
with
the deck.
1) Reagents cartridge contents are
thawed (leave at room temperature
for 45 minutes).
2) Reagents cartridge tubes are
uncapped and pressed into place at the
Reagents station so that the tubes are
firmly seated
and
level
with the deck.
3) Library Sample tube is
firmly seated
and in the correct orientation with
barcode facing the PCR Plate and
cap removed.
Chapter 4
Start the Ion Chef
™
run
Load the Ion Chef
™
System
4
36
Ion 520
™
& Ion 530
™
ExT Kit − Chef User Guide