Inspect and clean the Recovery centrifuges and buckets
1.
Inspect the Recovery centrifuges, then clean the components if excess liquid is present.
Is liquid present?
Action
No
Proceed to “Start the cleaning”
Yes
Clean the centrifuge bowl and buckets as described below.
IMPORTANT!
Clean the Recovery centrifuges occasionally, only when excess
liquid is noticeable in the bowl and/or buckets. You do
not
need to clean the
centrifuges after every run.
IMPORTANT!
Wear powder-free, nitrile gloves when cleaning the Recovery centrifuge.
2.
Remove the buckets from the Recovery centrifuges. Clean the inside and outside of each bucket
using a lint-free wipe, then place the buckets on a clean, dry surface while you clean the
centrifuge.
1
2
1
Bucket
2
Lint-free wipe
3.
Use lint-free wipes to remove all fluid from inside the centrifuge bowl.
1
2
1
Inside rim of the centrifuge
2
Bottom of the centrifuge bowls
4.
Use lint-free wipes treated with 70% isopropanol to clean the following parts.
• Inside rim of the centrifuge.
• Bottom of the centrifuge bowl.
• Outside and inside of the centrifuge buckets.
Chapter 7
Clean the Ion Chef
™
System
Clean the Ion Chef
™
Instrument
7
Ion 520
™
& Ion 530
™
ExT Kit − Chef User Guide
57