![Thermo Scientific Invitrogen Countess II FL Скачать руководство пользователя страница 63](http://html1.mh-extra.com/html/thermo-scientific/invitrogen-countess-ii-fl/invitrogen-countess-ii-fl_user-manual_1104588063.webp)
Product specifications
Technical specifications
Instrument type:
Benchtop cell counter and suspension cell-
based assay platform
Instrument dimensions:
9 (W) × 5½ (D) × 9 inches (H)
Weight:
8 lbs
Operating power:
100–240 VAC, 0.58 A MAX
Frequency:
50/60 Hz
Electrical input:
12 VDC, 2 A
Installation site:
Indoor use only, Class A Environments (i.e.,
non-residential or light industrial);
Pollution degree 2.
Operating temperature:
4°–40
℃
(39°–104°F)
Operating humidity:
<80% (non-condensing)
Processing time:
∼
15 seconds
Sample concentration range:
1 × 10
4
–1 × 10
7
cells/mL
Particle/cell diameter range:
4–60 µm (particles); 7–60 µm (cells)
Required sample volume:
10 µL
Firmware:
Countess
™
Automated Cell Counting
Platform Software
USB Drive :
4 Gigabytes
Optics:
3 channels (brightfield and 2 slots for
EVOS
™
LED light cubes)
Camera:
5 Mega pixels, 2.5× Optical Magnification
Material:
Poly(methyl methacrylate) (PMMA)
Dimensions:
25 mm (W) × 75 mm (D) × 1.7 mm (H)
Chamber volume:
10 µL
B
Physical
characteristics
Technical
specifications
Optics
Analysis slide
Countess
™
II FL automated cell counter User Guide
63
Содержание Invitrogen Countess II FL
Страница 81: ......