• Wipe the touch-screen of the Countess
™
II FL Automated Cell Counter using a
soft, lint-free cloth moistened with an LCD cleaning solution. Do not apply
excessive force during cleaning. Wipe the touch-screen dry immediately after
cleaning.
• Ensure that the cleaning solution does not enter the power button, the power
inlet, the slide port, or the USB ports.
• Never pour or spray any liquids directly on the instrument to avoid electrical
shock when the instrument is plugged in.
• Do not use abrasive cleaning solutions or material to prevent the touch-screen
from getting scratched.
• Wipe the instrument case of the Countess
™
II FL Automated Cell Counter using a
soft, lint-free cloth moistened with distilled water. Wipe the instrument dry
immediately after cleaning.
• Ensure that water or other cleaning solutions do not enter the power button, the
power inlet, the slide port, or the USB ports.
• Never pour or spray any liquids directly on the instrument to avoid electrical
shock when the instrument is plugged in.
• Wipe the instrument case of the Countess
™
II FL Automated Cell Counter using a
soft, lint-free cloth moistened with 70% alcohol. Wipe the instrument dry
immediately after cleaning.
• Avoid using a bleach solution, because it may leave a residue of bleach crystals
on the instrument.
• Ensure that water or other cleaning solutions do not enter the power button, the
power inlet, the slide port, or the USB ports.
• Never pour or spray any liquids directly on the instrument to avoid electrical
shock when the instrument is plugged in.
Set nominal focus
Nominal focus is the Z-point (i.e., depth) around which the auto focus function
searches to provide fine focus to the sample.
The auto focus algorithm of the Countess
™
II FL Automated Cell Counter is designed
to highlight the differences between live and dead cells in the brightfield channel. The
optimal focus level is where the live cells have a light colored center and the dead
cells are dark throughout (see examples below).
Clean the touch-
screen
Clean the
instrument case
Decontaminate
the instrument
Overview
Chapter 9
Maintenance
Set nominal focus
9
Countess
™
II FL automated cell counter User Guide
57