• To permanently save the results, press Save (Chapter 7, “Save results“).
• To perform a new count, remove the slide and reinsert it second chamber first
into the instrument, or insert a new sample slide.
View results
The Results screen for cell fluorescence assays displays a composite image of the
objects counted and the results of the cell count and cell viability calculations (total
concentration, percentage and concentration of cells counted through each
fluorescence channel).
Note:
The total cell concentration displayed after a fluorescent count does not take
any dilution into account. Therefore, the results reflect the actual cell concentration in
the sample slide, which must be multiplied by any dilution factor present to calculate
the original cell concentration.
This is in contrast to the cell counts in brightfield, where the counting algorithm
assumes a 1:1 dilution of the sample in trypan blue and displays the original cell
concentration (i.e., before the dilution) in the Results screen.
Identify objects counted
The Advanced
™
screen allows you to identify the objects (i.e., cells) counted in each
channel and included in the count results for further review. After reviewing the
marked objects, you can adjust the threshold for size, brightness, and/or circularity as
desired for your application.
Results screen for
cell fluorescence
assays
Advanced
™
screen
Chapter 5
Fluorescence assays
View results
5
34
Countess
™
II FL automated cell counter User Guide