Handling and Control
Programming
7-26
Heratherm IMP
Thermo Scientific
Programming
The menu item
Program
enables the user to create, store and launch up to 10 programs for
automating workflows. Each of the 10 programs may consist of a maximum of 10 steps.
The following properties may be defined for each step, depending on the current hardware
configuration:
• duration in hh:mm
• temperature set value
To ease the creation of programs users may choose to copy and edit existing programs.
End of Program
The end of a program is indicated by an audible signal (5 beeps). The refrigerated incubator is
switched off or changes to Temperature Hold mode, as preset in the preferences under
. The display shows the message
Program End
as well
as the temperature of the residual heat.
Instructions
Table 7-28 Launching a Program
Press
to activate the menu bar, then use
or
to select the
Program
icon and press
to
confirm.
In the selection screen shown at left, choose the
preselected option
Start
by pressing
.
Stop Timer
Prompt Upon Launching a Program
When a timer is running, an additional
Stop Timer
prompt
appears at this point (see
“Stopping an Off Timer Before It
Expires”
on
page
7-14
). Choosing Yes aborts the timer and
launches the selected program. Choosing No allows the
timer to continue running. However, when a user program
and a timer overlap, the program about to launch has priority
over the timer, that is, the timer will be put on hold and exe
-
cute only after the program has finished.
In the list of existing programs (for example:
P1...P4
)
shown at left, maintain the default selection
P1
or
choose another program using
(or
).
In either case, press
to launch the program.
2010-04-13 10:25
AM
Program
Start
2010-04-13 10:25
AM
Start
P1 P2 P3 P4
Eco
Содержание IMP 180
Страница 1: ......
Страница 6: ...Contents iv Heratherm IMP Thermo Scientific...
Страница 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Страница 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Страница 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Страница 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Страница 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Страница 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Страница 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Страница 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...