Thermo Scientific
Heratherm IMP
3-1
3
Installation
Ambient Conditions
Location Requirements
The refrigerated incubator must only be operated in a location that meets all of the ambient
condition requirements listed below:
• Installation location indoors in dry areas free from drafts.
• The dust burden may not exceed the contamination category 2 based on EN 61010-1.
Using the refrigerated incubator in an atmosphere with electrically conductive dust is
prohibited.
• The minimal distance to adjacent surfaces must be observed on all sides (see Section
• The operating room must be equipped with appropriate ventilation.
• Solid, level, fire-proof surface; no flammable materials opposite to the rear panel of the
refrigerated incubator.
• Vibration-proof substructure (floor stand, lab table) capable of bearing the dead weight of
the refrigerated incubator and its accessories (particularly if two devices are stacked).
• The electrical circuitry of the refrigerated incubator has been designed for an operating
height of up to 2000 m above sea level.
• If a high-voltage test is to be performed on the unit, it must first be heated for around 30
minutes at 50°C.
• The ambient temperature must be within a range of +18 °C to +28 °C (64 °F to 82 °F).
• Avoid direct exposure to sunlight.
• Devices that produce excessive amounts of heat must not be placed near the refrigerated
incubator.
• Place the refrigerated incubator on a floor stand (option; to be ordered separately), never
on the lab floor. This prevents the penetration of dust or dirt into the device.
• Power line voltage variations must not exceed ±10 % of the nominal voltage.
• Transient surges must lie within the range of levels that normally occur in the power
supply system. The impulse withstand voltage based on surge category II of
IEC 60364-4-443 shall be applied at the nominal voltage level.
Содержание IMP 180
Страница 1: ......
Страница 6: ...Contents iv Heratherm IMP Thermo Scientific...
Страница 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Страница 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Страница 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Страница 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Страница 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Страница 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Страница 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Страница 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...