Thermo Scientific
Heratherm IMP
4-1
4
Product Description
Heratherm IMP Series refrigerated incubator Overview
IMP Series advanced protocol microbiological refrigerated incubators come equipped with the
following features:
• high-precision work space temperature control, adjustable in steps of one-tenth of a degree
from 5°C (32°F) up to 70°C (158°F)
• Peltier-air-air-system with work space and outside fan
• countdown, fixed-time, and weekly timers for timed process control
• two perforated shelves
• an access port for tubing, sensor leads, etc.
Door frame heater
In Heratherm IMP series incubators frame heating in the inner door-box is installed to
prevent or reduce the risk of condensate formation.
Over-temperature protection (TWB)
The monitor temperature (maximum over temperature) can be set on the display manually.
When the maximum temperature is exceeded over the TWB the Peltier heating function is
switched off permanently and can only be reset by the user manually on the display again
(Instructions: "Settings" on page 7-21).
Содержание IMP 180
Страница 1: ......
Страница 6: ...Contents iv Heratherm IMP Thermo Scientific...
Страница 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Страница 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Страница 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Страница 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Страница 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Страница 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Страница 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Страница 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...