Product Description
Sensing and Control System
4-8
Heratherm IMP
Thermo Scientific
Figure 4-8 PT100 Sensor IMP 400
The work space temperature sensor provides the inputs to the refrigerated incubator’s built-in
controller, which continuously compares the measured values to the user-specified set value
and adjusts the heaters according to the result.
The unit has a thermal protection function that is factory-preprogrammed and not adjustable.
It protects the cultures in the work space from overheating: Thermal protection kicks in on a
brief violation of the upper limit, based on the defined setpoint temperature, at between 2 and
3 °C (35.6 °F and 37.4 °F) (37 °C (99 °F): 2 °C (35.6 °F), > 50 °C (122 °F): 3 °C (37.4 °F)),
automatically reducing the work space temperature to the user-specified set value and allowing
the incubation process to be continued even in case of a controller malfunction. If the thermal
protection is activated, the error message (E111) “Temperature too high” appears in the
display window and an audible alarm is sounded.
When the user acknowledges the error message, the red alarm icon (D4 in
figure 7-1
on
page 7-1
) is illuminated and the Temperature Set Value icon (see
table 7-3
on
page 7-4
) is
highlighted by a red border to indicate that thermal protection has kicked in.
Do not remove the protection hose from the sensor.
Protect the sensor from mechanical damage.
Содержание IMP 180
Страница 1: ......
Страница 6: ...Contents iv Heratherm IMP Thermo Scientific...
Страница 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Страница 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Страница 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Страница 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Страница 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Страница 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Страница 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Страница 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...