Installation
Installing the Anti-tilt Anchor
3-8
Heratherm IMP
Thermo Scientific
Installing the Anti-tilt Anchor
Table-top refrigerated incubators
The Heratherm table-top refrigerated incubators must be secured with anti-tilt anchors witch
are connected a solid part of a building. The anti-tilt anchor is to be mounted in the middle
position (2) or if not possible on the side opposite of the door hinges.
Bend the fixing tabs of the anti-tilt anchor up on one side and down on the other by an angle
of approx. 90°.
1. Do not use this position if the door is hinged on this side. Right-hand hinges represent the
standard configuration.
2. Preferred position.
3. Alternative position. Do not use if the door is hinged on this side.
Remove the bracket screws. Use the preferred position, if possible.
Fix the anti-tilt anchor with the bracket side down to the unit.
Position the unit with the anti-tilt anchor to in an angle of approx. 90° +/- 20%.
Take care that the stacking feet of the unit are still in correct place on the lower unit or on the
stacking adapter.
Fix the anti-tilt anchor to a solid part of the building.
Содержание IMP 180
Страница 1: ......
Страница 6: ...Contents iv Heratherm IMP Thermo Scientific...
Страница 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Страница 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Страница 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Страница 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Страница 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Страница 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Страница 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Страница 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...