Product Description
Data Communications & Alarm Interface
Thermo Scientific
Heratherm IMP
4-9
Data Communications & Alarm Interface
All signal connections are installed in the electrical interface panel at the rear of the
refrigerated incubator.
RS-232 Interface
The RS- 232 interface (item 2 in
below) may be used to connect Heratherm
refrigerated incubators to the serial interface port of a computer to allow for the
computer-aided acquisition and documentation of major operating parameters (temperature,
error codes, etc.).
[1]
Sample Sensor connection
[2]
RS 232 interface
[3]
Alarm contact
[4]
Power socket
[5]
2 fuses T 2A H 250 V, for socket
Figure 4-9 Signal Interfaces and Power Socket
Alarm Contact
The refrigerated incubator can be connected to an on-site, external alarm system (such as a
private branch telephone exchange, a facility monitoring system, visual or audible alarm
indicators). For this purpose, the devices come with a pre-wired potential-free alarm contact
(item 3 in
)
.
The alarm contact is energized whenever an error occurs in an internal
control loop or the refrigerated incubator’s electrical circuits or hardware.
AC Power Socket
The refrigerated incubator is connected to the AC supply mains via the socket (item 4 in
), which accepts a power cord with an IEC standard plug.
Fuses
15 A device fuses
Two 15 A slow-blow fuses mounted on the refrigerated incubator’s main electronic circuit
board protect internal circuitry from the impact of excessive power consumption.
Содержание IMP 180
Страница 1: ......
Страница 6: ...Contents iv Heratherm IMP Thermo Scientific...
Страница 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Страница 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Страница 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Страница 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Страница 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Страница 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Страница 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Страница 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...