Maintenance
Replacing the Door Seal
Thermo Scientific
Heratherm IMP
10-5
Replacing the Door Seal
The door seal of the outer door is located in the retaining slot.
The door seal should be inspected for any signs of embrittlement at half-yearly intervals.
No tools are required to replace the seal.
Figure 10-1 Door Seal Replacement
1. Pull the seal out of the guide slot.
2. Starting on the hinge side of the door, position the end of the new seal at the location
above.
3. Gently press the seal into the slot, working around the circumference of the door. In
corner areas in particular ensure that the seal lip is installed without any wrinkles and that
the seal is not stretched or compressed.
4. Make sure that the seal is positioned correctly in the slot and flush with the door frame.
Correct, if necessary.
Special door seals
There are special door seals available, please refer to the
accessory list:
• Viton gasket made for silicone-free applications
• Special gasket for higher humidity applications (reduces risk
of condensation)
Содержание IMP 180
Страница 1: ......
Страница 6: ...Contents iv Heratherm IMP Thermo Scientific...
Страница 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Страница 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Страница 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Страница 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Страница 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Страница 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Страница 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Страница 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...