Installation
Transport
3-6
Heratherm IMP
Thermo Scientific
Floor stand refrigerated incubators
The floor stand refrigerated incubators come equipped with four (4) casters. The lever for
releasing the caster is located above the locking lever. After positioning the unit in its
installation location ensure that the locking levers are pressed down on the casters.
To ensure the degree of stability specified by safety requirements the front casters must be
turned so that they are facing forward after the unit has been positioned in its installation
location and the locking levers pressed down on these casters.
Danger of tipping when moving!
Disconnect the unit from the power source before moving it.
Unscrew the wall mounts/supports from the wall.
Retract the spacers.
Move the Heratherm floor stand units with caution.
When moving the unit always give particular attention to protruding items,
such as door handle, spacers, etc.
Quick starts and stops can result in tipping!
Always ensure that the doors are closed when moving the unit.
Содержание IMP 180
Страница 1: ......
Страница 6: ...Contents iv Heratherm IMP Thermo Scientific...
Страница 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Страница 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Страница 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Страница 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Страница 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Страница 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Страница 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Страница 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...