Thermo Scientific
Heratherm IMP
6-1
6
Operation
Preparing the refrigerated incubator
The refrigerated incubator must not be released for operation before all major start-up
activities have been completed (see
“Installation procedures”
on
page 5-1
).
Device Check
Prior to starting operation, the following refrigerated incubator components must be checked
for their correct function:
• The door seal in the front frame must not be damaged.
• The glass door must not be damaged.
• The shelving components must be installed safely.
• Disinfecting the refrigerated incubator’s Work Space
Disinfect the work space according to the operator-specified hygiene guidelines.
Starting Operation
1. Turn the refrigerated incubator on using the control panel.
2. Adjust the temperature set value on the control panel.
3. The temperature controller starts adjusting the work space to the user-specified
temperature set value now.
4. Load the work space with samples.
To avoid any risk of explosion or fire
• refrain from loading the refrigerated incubator with any of the
substances listed in the section
“Incorrect Use”
on
page
1-6
• make sure that the ambient air is free of any solvents
• do not operate the refrigerated incubator in areas with an
explosion hazard
Содержание IMP 180
Страница 1: ......
Страница 6: ...Contents iv Heratherm IMP Thermo Scientific...
Страница 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Страница 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Страница 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Страница 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Страница 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Страница 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Страница 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Страница 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...