Installation procedures
Connecting the RS-232 Interface
Thermo Scientific
Heratherm IMP
5-9
Users may employ the RS-232 command inventory listed in
below for automating
process data logging - for example, by embedding these commands in scripts that run on a
remote computer.
Interconnecting the refrigerated incubator with a Computer
1. Turn the computer off.
2. Route the serial interface cable along a path that does not cross hot exhaust air piping,
Peltier elements, stacked devices, tables, aisles or passageways.
With stacked devices, keep the serial interface cable away from hot spots on the other
refrigerated incubator in the stack.
3. Connect one connector of the serial interface cable (cable length, 5 to max. 10 m, not
supplied as a standard item) to the socket labeled
RS 223
in the computer and alarm
interface section at the rear of the refrigerated incubator (see
“Signal Interfaces and Power
Socket”
on
page 4-13
).
4. Connect the second connector to an unused COM 1 /COM 2 or other serial port on the
computer.
5. Boot the computer.
6. Launch your standard terminal program and set up the connection with the following
parameters:
— 57600 bits per second
— 8 data bits
— 1 stop bit
— No Parity
7. Once your terminal indicates that serial communication has been established successfully,
enter any of the commands listed in
below, depending on what type of
information you want to query.
8. Use the following generic command syntax:
?:aaaa:bb::cc<CR>
, where:
—
?:
identifies the command line as a query;
—
aaaa:
is the parameter address;
RS-232 interface compatibility
To avoid overloading and damaging the RS-232 interface check the
interfacing parameters against the pin-out description given above and
make sure that computer’s interface port works with a signal level of +/-
5V
DC.
Содержание IMP 180
Страница 1: ......
Страница 6: ...Contents iv Heratherm IMP Thermo Scientific...
Страница 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Страница 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Страница 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Страница 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Страница 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Страница 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Страница 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Страница 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...