Product Description
Safety Devices
Thermo Scientific
Heratherm IMP
4-7
Safety Devices
The refrigerated incubators are equipped with the following safety features:
• a sample protection feature that safeguards the samples against destruction through
overheating in case of controller failure;
• dual sensor: if one sensor fails, the other will take over.
• an overheat protection cut-out feature class 2 (TWB), according to DIN 12880, that shuts
down the refrigerated incubators IMP completely when excessive temperatures occur in
the workspace;
• an alarm relay that is energized, along with audible and visual alarms, to indicate errors
encountered during operation;
• dual fuses rated at 15 amperes.
Work Space Atmosphere
To ensure undisturbed operation, the ambient temperature in the operating room must be at
least 18 °C (64.4 °F).
The Peltier element controls the work space temperature of the refrigerated incubator from
5 °C (41 °F) up to the maximum of 70 °C (158 °F).
Sensing and Control System
The PT 100-type sensor for the control of the work space temperature and for the thermal
protection is installed on the bottom of table-top refrigerated incubators and in the top of
floor stand refrigerated incubators.
Figure 4-7 PT100 Sensor IMP 180
Содержание IMP 180
Страница 1: ......
Страница 6: ...Contents iv Heratherm IMP Thermo Scientific...
Страница 8: ...List of Figures ii Heratherm IMP Thermo Scientific...
Страница 40: ...Product Description Additional Tube Access Ports Option 4 14 Heratherm IMP Thermo Scientific...
Страница 88: ...Handling and Control Programming 7 34 Heratherm IMP Thermo Scientific...
Страница 90: ...Shut down Shutting the refrigerated incubator down 8 2 Heratherm IMP Thermo Scientific...
Страница 110: ...Technical Data 13 4 Heratherm IMP Thermo Scientific...
Страница 114: ...Device Log 15 2 Heratherm IMP Thermo Scientific...
Страница 117: ...Contact Thermo Scientific Heratherm IMP 16 3...
Страница 118: ...Contact 16 4 Heratherm IMP Thermo Scientific...