Consumables
Cooling Water
5-4
iCAP RQ Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
Thermo Scientific
lists the technical specifications for the NESLAB
ThermoFlex 2500 recirculation chiller, which is optionally shipped with
the iCAP RQ instrument.
NOTICE
For information on the power supply of the chiller, see
▲
Table 5-2.
Chiller specifications
Chiller Component
Specification
Cooling Capacity
(at 20 °C ambient temperature)
2.2 kW (50 Hz); 2.5 kW (60 Hz)
Ambient Temperature
10 to 40 °C
Water Temperature
5 to 40 °C
Stability of Temperature
Regulation
±0.1 °C
Reservoir Volume
7.2 L
Flow Rate
Turbine Pump (60 Hz)
Turbine Pump (50 Hz)
13.3 L/min at 0.41 MPa (4.1 bar)
9.5 L/min at 0.41 MPa (4.1 bar)
Unit Dimensions (H × W × L)
73.6 cm × 43.6 cm × 67.3 cm
Unit Weight
79.6 kg