Site Preparation
Dimensions and Weights
Thermo Scientific
iCAP RQ Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
2-5
2. With a pallet jack, lift the instrument to the height of the working
bench.
3. Remove the four clamps (see
) that fix the instrument to
the transport pallet.
4. Hold the lifting device so that the yellow label on the lifting device
which advises users to read the manual is visible. Place the lifting
device on top of the handles, with the two ends of the twice bent
tube pointing toward the instrument.
5. Position the correct holes in the metal sheet of the lifting device
above the handles of the instrument, see
.
Each lifting device consists of a folded metal sheet with holes and a
twice bent tube. It is delivered with four hexagon screws, four
washers, and two clamps.
6. To mount a lifting device onto the handles at one instrument side,
place a clamp (
3
) below each of the two handles.
Figure 2-3.
Positioning the lifting device above the handle of iCAP RQ
instrument
left side of the
instrument
(seen from the front)
right side of the
instrument
(seen from the front)