vii
iCAP RQ ICP-MS Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
Thermo Scientific
Plasma and Cooling Gas
Argon
Purity
99.996% or better
Max. water content
5 ppmv
Supply rate
max. 24 L/min
Pressure
min. 0.55 MPa (5.5 bar)/max. 0.6 MPa (6 bar)
CCT Gas
Helium (for QCell)
Purity
99.999% or better
Max. water content
2 ppmv
Supply rate
max. 10 mL/min per channel
Pressure
min. 0.05 MPa (0.5 bar)/max. 0.15 MPa (1.5 bar)
Hydrogen mixtures with helium
(for QCell)
Max. content of hydrogen
4.5%
Purity
99.999% or better
Max. water content
2 ppmv
Supply rate
max. 10 mL/min per channel
Pressure
min. 0.05 MPa (0.5 bar)/max. 0.15 MPa (1.5 bar)
Oxygen or mixtures with helium
(for QCell)
Purity
99.999% or better
Max. water content
2 ppmv
Supply rate
max. 10 mL/min per channel
Pressure
min. 0.05 MPa (0.5 bar)/max. 0.15 MPa (1.5 bar)
Ammonia or mixtures with helium
(for QCell)
Max. content of ammonia
1%
Purity
99.999% or better
Max. water content
2 ppmv
Supply rate
max. 10 mL/min per channel
Pressure
min. 0.05 MPa (0.5 bar)/max. 0.15 MPa (1.5 bar)
Additional Gas
Argon
Purity
99.996% or better
Max. water content
5 ppmv
Supply rate
max. 1 L/min per channel
Pressure max.
0.1 MPa
(1 bar)
Pure helium or mixtures with
argon
Purity
99.996% or better
Max. water content
5 ppmv
Supply rate
max. 1 L/min per channel
Pressure max.
0.1 MPa
(1 bar)
Pure oxygen or mixtures with
helium and/or argon
Purity
99.996% or better
Max. water content
5 ppmv
Supply rate
max. 1 L/min per channel
Pressure max.
0.1 MPa
(1 bar)
Pure nitrogen or mixtures with
helium and/or argon
Purity
99.996% or better
Max. water content
5 ppmv
Supply rate
max. 1 L/min per channel
Pressure max.
0.1 MPa
(1 bar)
Methane mixtures with argon
Max. content of methane
4.5%
Purity
99.996% or better
Max. water content
5 ppmv
Supply rate
max. 1 L/min per channel
Pressure max.
0.1 MPa
(1 bar)