Site Preparation
Placing the System
Thermo Scientific
iCAP RQ Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
2-11
Floor Space for the System
shows an example of the required floor space for the system.
All values are given in millimeter.
Placing the Fore Vacuum Pump
The iCAP RQ ICP-MS instrument is delivered with a fore vacuum
pump (Sogevac™ SV40BI), a data cable and a grounding cable for
connecting the instrument to the fore vacuum pump, a power cable for
connecting the pump to the mains, and an exhaust hose for connecting
the fore vacuum pump to the exhaust system (instrument). A noise
reduction cover and a drip pan for the fore vacuum pump are available
as options.
The fore vacuum pump must be placed on a drip pan of sufficient
volume. Order a drip pan at Thermo Fisher Scientific or use a suitable
one of your own.
NOTICE
Thermo Fisher Scientific recommends the use of the noise
reduction cover when installing the fore vacuum pump.
▲
Labeled Components: 1=monitor, 2=mouse, 3=keyboard, 4=standard workbench, 5=iCAP RQ MS, 6=wall
Figure 2-9.
Required floor space for iCAP RQ instrument
700 - 1000
1200
748
767
15
1000
5
6
4
1
2
3
772