Operating Environment
Vibration
Thermo Scientific
iCAP RQ Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
3-7
Vibration
Floors must be free of vibration caused, for example, by equipment in
adjoining locations. Propagation of vibrations and their influence on
complex instrumentations are difficult to predict. Contact your local
Thermo Fisher Scientific representative if you have questions or
concerns about your laboratory.
NOTICE
The fore vacuum pump must not have any mechanical contact
to the mass spectrometer with exception of the vacuum hose during
operation. The vibration of the fore vacuum pump might impede the
performance of the instrument. Install the fore vacuum pump on the
floor beneath the mass spectrometer. Do not install the fore vacuum
pump near the system on the workbench.
▲
Airborne Noise Emission
The A-weighted emission sound pressure level created by the iCAP RQ
mass spectrometer and fore vacuum pump at work stations does not
exceed 70 dB(A).
Radio Frequencies
The iCAP RQ system withstands electric fields of 1 V/m and a
frequency range of 80 MHz to 2.7 GHz without any influence to
operation.
iCAP RQ instruments are designed to work in a controlled
electromagnetic environment. Do not use radio frequency transmitters,
such as mobile phones, in close proximity to the instrument.
If strong radio transmitters are operating close to your laboratory,
contact your local Thermo Fisher Scientific office for advice.
Magnetic Fields
The instrument site must be free of interfering magnetic fields. The
maximum acceptable field amplitude (AC) for any frequency is 5.5 mT.
NOTICE
Sources of disturbing fields are, for example, other analytical
instruments such as NMR systems or Zeeman AAS, train, tram, subway,
high power cables crossing the ceiling, large electric motors (elevators),
radio stations nearby.
▲