Site Preparation
Entrance Requirements
2-2
iCAP RQ Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
Thermo Scientific
Entrance Requirements
To allow moving a packed iCAP RQ ICP-MS instrument, the entrance
to your facility and the width of all hallways, elevators, etc., should have
a minimum width of 89 cm (35.1 in.).
Thermo Fisher Scientific recommends checking whether the container
with the instrument fits through the laboratory entrance. Also allow
additional room for maneuvering the system around corners, into
elevators, or through doorways. Note that it is necessary to use a means
of transport (a pallet jack, for example).
NOTICE
Do not remove the instrument from its shipping container
unless authorized by Thermo Fisher Scientific personnel. Be sure that all
the contents of the container remain with the instrument.
▲
Dimensions and Weights
The iCAP RQ ICP-MS instrument is shipped in a container. Other
modules such as the fore vacuum pump are shipped in separate
containers. For dimensions and weights, see
NOTICE
Owing to the climatic conditions in some tropic regions, some
boxes may be replaced by special packings. As a result, the dimensions
will differ from those given in
.
▲
Additional weights and the dimensions of auxiliary boxes might be
added, depending on the equipment ordered.
The floor of your laboratory must be able to accommodate the weight of
the instrument including all components and any options that are added
to the system. See also
shows the external dimensions of the system. All values are
given in millimeter.
Table 2-1.
Packing information of a typical iCAP RQ instrument
Module
Weight Gross/Net [kg]
Dimensions
L × W × H [cm]
Basic Unit
195/142
133 x 90 x 160
Palette with 8 containers
(fore vacuum pump,
installation kit, computer,
pump oil, etc.)
103/89
120 x 80 x 82
Total Weight 298/231
Storage Space 2.3 m
³