Site Preparation
Placing the System
2-10
iCAP RQ Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
Thermo Scientific
Placing the System
Workbench for Instrument
The workbench for the iCAP RQ system must be capable of supporting
the weight of the mass spectrometer (
142
kg) plus the weight of any
option (autosampler, liquid chromatograph, for example) and stand in a
secure and level position.
The iCAP RQ instrument is designed to be placed on a bench with its
rear panel against a wall. The bench surface should ideally be 80 to
85 cm above the floor.
On the left side panel, free access to the mains switch and circuit breaker
is needed to allow shutting off the instrument in an emergency at all
times. To allow shutting off the instrument in an emergency, free access
must be possible at any time to the main power switch and the power
connections at the right and left side of the instrument.
Allow at least 70 to 100 cm of clear space left of the system for clearance
of the gas lines, electrical connections, as well as for the exhaust line and
vacuum hose of the fore vacuum pump.
It is recommended to leave sufficient space on the right side of the
iCAP RQ instrument for the connections to the computer and
peripheral devices.
Consider suitable space between the outputs of the fume extraction lines
and the inputs on the instrument. Consider the total height of the
system when it is located on the table.
!
Use only workbenches that are able to carry the weight of the iCAP RQ
instrument and provide sufficient stability. The workbench must be
dry and clean.