Line Power
Electrical Power Requirements
4-4
iCAP RQ Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
Thermo Scientific
Auxiliary Wall Outlets
Additional single-phase outlets are needed for additional parts such as
fore vacuum pump, computer and monitor. Thermo Fisher Scientific
recommends at least five spare outlets in the near vicinity of the left side
of the system and five close to the workbench space within the your
laboratory.
The components of the data system (computer, monitor) require wall
outlets at a nominal voltage of 200-240 V AC, 50/60 Hz, also see the
manufacturer’s manual for more information.
The fore vacuum pump requires a separately fused wall outlet due to the
high current of the pump. This wall outlet must be fused with 15 A or
16 A, tripping characteristic ‘C’ or ‘D’.
For information on the power outlet requirements, also see the
manufacturer’s manuals.
Fore Vacuum Pump
The fore vacuum pump requires a dedicated power outlet that complies
with the specifications as indicated on the rating plate of the fore
vacuum pump, see
Quality of Power
The iCAP RQ instrument complies with the requirements listed in
Chapter 6: “General Pre-Installation Information”
details on the required quality of power.
Table 4-2.
Power requirements for fore vacuum pump
Voltage
Frequency
Power
200-240 V ±10%
50 Hz
1.6 kW
200-240 V ±10%
60 Hz
1.9 kW
Table 4-3.
Immunity compliance of iCAP RQ ICP-MS
Description
Requirement
Immunity to Electrical Fast Transient/Burst
EN 61326-1, EN 61000-4-4
Immunity to Electrical Slow Transient/Surge
EN 61326-1, EN 61000-4-5
Immunity to Conducted RF Voltage
EN 61326-1, EN 61000-4-6
Immunity to Voltage Dips, Short Interruptions and Voltage
Variation
EN 61326-1, EN 61000-4-11