1500 Series A2 1-9
Thermo Scientific
Section 1
Description
The optional, factory-
installed UV light is
installed at the top of the
front section of the sample
chamber.
The operating time of the
UV light is preset for 1
hour. The UV disinfection
procedure can be started by
pressing the “UV” key on
the control panel with the
window in the closed
position. If the window is
opened, the procedure is
immediately cancelled.
Note
To protect from UV radiation, the UV light can only be turned on
when the window is completely closed.
s
The single piece work tray is provided with the cabinet. It is placed onto
the frame above the sample chamber drain pan. It can be removed by
lifting up, using the large holes in the work tray.
The working area for optimal product protection extends over the entire
width and depth of the work tray. The two armrests are positioned in the
center of the working area 8 inches (20 cm) from each other. To attach
the armrests, insert into the first perforated track of the work tray.
Warning
Work safety is assured only if the armrests are used correctly!
s
Work Area
UV Lights
UV light
Chamber light
Figure 1-7.
Lights
Width
Depth
Perforation track
Armrest
Holes for padded
armrests (Optional)
Armrest
Separation
distance
Perforation
track
Work tray
Figure 1-8.
Work tray