Tapping the “UV Disinfection” button takes you to the “Disinfection
Settings” screen, shown in Figure 4-31 below. From this screen, you can
change the default running time of the UV disinfection cycle. This value is
used by default when you start the Disinfection cycle by tapping the “UV
light” button in the Control Panel Area of the display window.
Figure 4-31.
“Disinfection Settings” Screen
The “Disinfection Cycle Settings” screen displays the following fields:
• “UV Cycle Duration”: Sets the duration of the UV Disinfection cycle
from 0 to 23 hours, with additional intervals of 00, 15, 30, and 45
minutes. Hours are entered by using a calculator-style keypad, minutes
are selectable from a drop-down list. After UV cycle has been
completed, the default setting is to enter standby mode, and you can
also choose to enter idle mode by selecting the “Idle mode after UV
cycle has been completed”
NOTE:
The default setting is 1 hour, 00 minutes. 0 hours and 00
minutes is NOT a selectable option.
• “UV Auto Start” & “UV Auto Start Time”: Enable/Disable the Auto
UV disinfection function by "UV Auto Start". If the Auto UV
disinfection function is enabled, you can also set the auto start time by
"UV Auto Start Time".
• Tap the “Save” button to save the setting.
1500 Series A2 4-21
Thermo Scientific
Section 4
Operation
UV Disinfection