The exhaust thimble connection is installed to the top of the cabinet using
an additional adapter. The threaded inserts for the retaining screws of the
adapter are pre-installed to the top of the cabinet.
Installation on a 6 ft. model:
1. Turn the unit off and disconnect it from the power source.
2. Clean the contact surfaces around the exhaust opening, and the upper
and lower contact surfaces of the adapter thoroughly to ensure that all
grease and dust residue are removed.
3. Secure the adapter to the pre-installed threaded inserts on top of the
cabinet using the supplied four retaining screws. Tighten the screws
hand-tight.
4. Secure the accessory by tightening the supplied retaining screws (M 5)
wrench-tight.
5. Using the adjustment nuts
on the sliding panel, set
the gap height to an
opening of 2 inches.
1. Place the connecting
sleeve onto the connecting
opening of the accessory
and secure it with the
screws.
2. Fit the connecting pipe
(200 mm [8 in]) of the
on-site exhaust air system
to the connecting sleeve.
3. Seal the unused
connecting opening with
the cover. Place the cover
onto the connecting
opening at the housing
and secure it with the
screws.
1500 Series A2 2-3
Thermo Scientific
Section 2
Installation
Connecting Sleeve
Exhaust System
Accessories (continued)
Thimble
Retaining
screws
Exhaust
opening
Threaded
inserts
Pipe
Sleeve
Cover
Sleeve
Opening
Adapter
Figure 2-2.
6 ft Thimble and Sleeve