background image

Caution

 Service and certification must be performed by qualified 

personnel. 

s

 

Biological safety cabinet certification consists of a series of tests designed to 
verify that the cabinet is performing within operating parameters 
established by the manufacturer. To assure that a biological safety cabinet 
is operating as intended, each cabinet should be field-tested at the time of 
installation and at least annually thereafter. Cabinets should be re-certified 
whenever HEPA filters are changed, internal maintenance is performed, or 
the unit is relocated.  

Three industry-related organizations maintain lists of companies and 
individuals who are active in the certification industry. You may contact 
these organizations at the addresses listed below. 

NSF International (NSF) and International Air Filtration Certifiers 
Association (IAFCA) sponsor certifier accreditation programs. Accredited 
certifiers have demonstrated proficiency at testing biological safety cabinets 
by successfully completing written and/or practical examinations. 

Biohazard Cabinet Field Certifier Program 
NSF International 
PO Box 130140 
789 N. Dixboro Rd 
Ann Arbor, MI 48113-0140 
Telephone (734) 769-8010 Or (800) NSF-MARK 
Fax (734) 769-0109 
http://www.nsf.org/Certified/Biohazard-Certifier 
 
IAFCA 
PO Box 12155 
Columbus, OH 43212 
Telephone (888) 679-1904 
Fax (614) 486-1108 
http://www.iafca.com/certifier.html

3-6         1500 Series A2

Thermo Scientific

Section 3

 

Start-Up 

Locating a Certifier

Содержание 1500 A2 Series

Страница 1: ...Visit us online to register your warranty www thermoscientific com labwarranty 1500 Series A2 Class II Type A2 Biological Safety Cabinet Operating Manual Rev 00 User Manual ...

Страница 2: ...atures that will enhance your safety and improve overall operating efficiency Should you have any questions on using this product or need further explanation of any of it s features please contact Technical Services Page iv or your local distributor Stainless Steel Interior Coated Interior Size Voltage 10 inch opening 8 inch opening 10 inch opening 8 inch opening 1574 1584 4 ft 230V China 1579 158...

Страница 3: ...rvice personnel s Material in this manual is for information purposes only The contents and the product it describes are subject to change without notice Thermo Fisher Scientific makes no representations or warranties with respect to this manual In no event shall Thermo Fisher Scientific be held liable for any damages direct or incidental arising out of or related to the use of this manual 2020 Th...

Страница 4: ...ry Hot surface s present which may cause burns to unprotected skin or to materials which may be damaged by elevated temperatures WEEE Compliance Thermo Fisher Scientific has contracted with companies for recycling disposal in each EU Member State For further information send an email to weee recycle thermofisher com 4 Always use the proper protective equipment clothing gloves goggles etc 4 Always ...

Страница 5: ...ipment We can fill your needs for spare or replacement parts or provide you with on site service We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products Whatever Thermo Scientific products you need or use we will be happy to discuss your applications If you are experiencing technical problems working together we will help you locate the problem and cha...

Страница 6: ...Connections 2 5 Universal Piping Connections 2 7 SmartPort 2 8 Drain Valve 2 8 Start Up 3 1 Assemble the Stand 3 1 Unlock Counterweight 3 3 Level the Cabinet 3 3 Power Connection 3 4 Installation Tests 3 5 Locating a Certifier 3 6 Operation 4 1 Initial Setup 4 1 Operation 4 3 Startup Mode 4 8 Working Mode 4 9 Standby Mode 4 9 Idle Mode 4 10 Running UV Disinfection 4 11 Viewing Status and Replacing...

Страница 7: ...rior Surfaces 5 4 Clean Window 5 4 Clean the Drain Pan 5 5 Clean the Paper Catch Grid 5 6 Maintenance 6 1 Service 6 2 Sample Chamber Lights 6 2 Optional UV Lights 6 2 Replacements and Repairs 6 3 Unit Disposal 6 3 Specifications 7 1 Certification Testing 8 1 Classification of the Safety Cabinet 8 1 Test Terms 8 1 Testing 8 2 Test Equipment 8 3 Testing Information 8 3 HEPA Filter Leak Test 8 8 Filt...

Страница 8: ...nditions even between annual certifications The plenum assembly consists of a plenum for the downflow blower and a plenum for the exhaust air blower Each blower includes a filter The exhaust air is discharged through an opening in the top of the cabinet Chamber lighting includes one LED bulb for 4 foot models and two for 6 foot models The factory installed UV light is installed in the ceiling at t...

Страница 9: ... plenum hose is located on the left and the exhaust plenum hose on the right Note Do not remove the caps of the two test hoses to check for supply and exhaust airflow s Negative pressure air system A negative pressure air system combines with HEPA filters in the supply and exhaust airflow for personnel and product protection Personnel protection Inflow air along the entire working opening at a con...

Страница 10: ...flow air velocities guarantees that product and personnel protection remain uncompromised Airflow monitoring determines the velocity of the airflow in the sample chamber as well as the inflow velocity of air through the exhaust opening As soon as airflow velocities rise above or fall below a specified safety value an audible and a visual alarm is activated Window position monitoring The position s...

Страница 11: ...e unit and exhausted as ultra pure air into the room environment or into an external exhaust system via an exhaust transition Sectional paper catch grids are installed beneath the work tray in the air duct between the inner and outer back wall of the unit These grids prevent large items such as paper towels and tissues from entering the plenum where they may impair the function of the blowers or t...

Страница 12: ... Set Back mode by closing the window When the window is closed the blower speed is reduced minimizing energy consumption and noise emittance without compromising particulate containment inside the work chamber Note The working and loading window positions are indicated on the user interface as controlled by micro switches and displayed by status indicators on the control panel s Use of the Window ...

Страница 13: ...ow in the closed or reduced mode position move it to the STANDBY position Figure 1 5 and Make sure that the window touches the gasket below Use of the Window continued 1 6 1500 Series A2 Thermo Scientific Section 1 Description Figure 1 4 Work Position Figure 1 5 Fully Closed Position UV if applicable dimple ...

Страница 14: ...op near the front of the cabinet The accessory outlets max 5A are located on the interior chamber back wall On top of the cabinet are the two fuseholders for the 5A accessory outlet fuses and an USB connection signal source not customer interface Service valve access ports are located on each side of the cabinet To use these ports remove the center perforation Service valves are available as optio...

Страница 15: ...ide walls resist glare and make it easier for the user to work safely Lighting for the sample chamber includes one LED bulb for 4 foot units and two bulbs for 6 foot units The bulb s is installed behind the light canopy Service valve Access ports Power cord USB Fuses N L Accessory outlet 1 of 2 shown Figure 1 6 Interfaces 2 of 4 access ports are shown Chamber Lighting Unit Interface continued ...

Страница 16: ... work tray is provided with the cabinet It is placed onto the frame above the sample chamber drain pan It can be removed by lifting up using the large holes in the work tray The working area for optimal product protection extends over the entire width and depth of the work tray The two armrests are positioned in the center of the working area 8 inches 20 cm from each other To attach the armrests i...

Страница 17: ......

Страница 18: ...at are not connected to an exhaust system the distance between the exhaust air opening at the device and the room ceiling must be at least 8 inches 200 mm Equipped with an appropriate ventilation system Room temperature between 10 C and 40 C 50 F and 104 F Relative humidity must not exceed 80 up to 31 C and decreasing linearly to 50 at 40 C For indoor use only Room ventilation should preferably be...

Страница 19: ... are installed to the exhaust opening on top of the cabinet The threaded inserts for the retaining screws are pre installed in the unit Installation on 4 model 1 Turn the unit off and disconnect it from the power source 2 Thoroughly clean the surfaces around the exhaust opening and of the accessory to be installed to ensure that they are free of debris 3 Secure the accessory by tightening the supp...

Страница 20: ...hten the screws hand tight 4 Secure the accessory by tightening the supplied retaining screws M 5 wrench tight 5 Using the adjustment nuts on the sliding panel set the gap height to an opening of 2 inches 1 Place the connecting sleeve onto the connecting opening of the accessory and secure it with the screws 2 Fit the connecting pipe 200 mm 8 in of the on site exhaust air system to the connecting ...

Страница 21: ...a move within a building Caution When moving the unit lift only at the lift points shown in Figure 2 3 s Caution Do not allow the weight of the cabinet to rest on the drain pan s Warning The weight of the window is balanced by the counterweight on the back of the cabinet Do not move the unit unless the counterweight has been locked in place with the shipping screws see Figure 3 2 s Figure 2 3 Lift...

Страница 22: ...emove the centers 4 Referring to Figure 2 5 on the following page slide the bezel onto the threaded pipe of the service valve Apply a thin bead of silicon to the inner flange of the service valve 5 From inside the sample chamber slide the service valve through the desired plug 6 From the outside slide the washer onto the service valve threaded pipe and apply a thin bead of silicon to the outer fla...

Страница 23: ... Washer Nut Union nut Figure 2 5 Valve Assembly and Installation If SmartPorts are installed replace the SmartPort grommet before starting a decontamination procedure Two replacements P N 114111 are shipped with each unit or the 1910112 kit containing 4 grommets can be ordered SmartPort locations are indicated above one on each side Replace the SmartPort Before Decontamination ...

Страница 24: ... Refer to illustration below Note The top left access point is not available on 3 cu ft units due to the proximity of the exhaust filter On 3 cu ft units the right side access points are next to the right side rope channel The left side access points are to the left of these s 1500 Series A2 2 7 Thermo Scientific Section 2 Installation Universal Piping Connections Figure 2 6 Possible Universal Pip...

Страница 25: ...s ensures that no particles escape into the lab At any time new solid rubber grommets may be installed into the SmartPort to provide a complete seal in the side walls It is recommended that new solid grommets be installed on the exterior side walls prior to decontaminating the cabinet Two new solid grommets arre supplied with the cabinet for this purpose or kit 1910112 four grommets can be ordered...

Страница 26: ...eveler into place 2 Locate the left and right side panels Install the telescoping leg assemblies into the side panels as shown in Figure 3 1 3 The height adjustable stand can be set to a height range of 26 8 34 7 in 680 880 mm To set the stand to the desired height pull the legs downward out of the channels to the desired height Lock the height adjustment in place by installing two lockwashers scr...

Страница 27: ...nnel Lockwasher Lockwasher Crossmember Retaining tab Screws Side Back Lockwashers Figure 3 1 Stand Assembly Assemble the Stand continued 6 Place the safety cabinet onto the stand so that the Allen screws are routed through the holes of the keyhole tabs 7 Slide the cabinet frame to the keyhole stop see Figure 3 1 8 Tighten the four Allen screws 9 Level the cabinet instructions follow in this sectio...

Страница 28: ...bag for this purpose Warning One person should hold the counterweight while a second person removes the screws This facilitates screw removal and avoids the possibility of stripping the screws s Note The cabinet should be leveled only after it has been located in the desired working location s For cabinets without a stand place a bubble level onto the work tray surface and adjust the cabinet until...

Страница 29: ... the receptacle does not match the power plug consult a qualified electrician to replace or install the correct receptacle 3 Make sure that the power cord is routed away from the counterweight and cable guide The left side of the cabinet has additional fastening points that can be used for routing the power cord 4 Make sure that the power cord is not stretched or pinched To protect from accidental...

Страница 30: ...ort from the authorized service technician Warning The operational safety of the unit particularly the personnel and product protection are guaranteed only if all safety functions of the unit have been tested and approved s Caution Thermo will not warrant operational safety if the unit is operated without the required installation and certification test or if these tests and repeat test are not pe...

Страница 31: ... and individuals who are active in the certification industry You may contact these organizations at the addresses listed below NSF International NSF and International Air Filtration Certifiers Association IAFCA sponsor certifier accreditation programs Accredited certifiers have demonstrated proficiency at testing biological safety cabinets by successfully completing written and or practical exami...

Страница 32: ...ation Controlled Environment Testing Association 1500 Sunday Drive Suite 102 Raleigh NC 27607 Telephone 919 787 5181 Fax 919 787 4916 http www cetainternational org members corp_indiv htm Note Unless certification was expressly called for in the specification quotes and or purchase order the cost for this on site testing is to be paid for by the customer s 1500 Series A2 3 7 Thermo Scientific Sect...

Страница 33: ......

Страница 34: ...rocess before you can use the unit Proceed as follows to start up the unit 1 Power up the device by connecting the external power source A startup screen with a Start Setup button appears prompting you to run the setup as shown below 2 Tap the Start Setup button The initial setup screen appears as shown below prompting you to choose a display language set the date and time Figure 4 1 Initial Start...

Страница 35: ...n MM DD YYYY to open the keypad to set the date 4 Tap the time button HH MM to open the keypad to set the time 5 Tap the Finish button to complete the initial set up The unit comes up in idle mode as shown below assuming that the window is closed Figure 4 3 Date Setup Figure 4 4 Time Setup Figure 4 5 Main screen Unit in Idle Mode Initial Setup continued ...

Страница 36: ...17 8 cm 7 inches diagonally across and has touch sensitive areas that support finger gestures like tapping it can be used with gloves The touchscreen display unit of the Biological Safety Cabinets is subdivided into the four main screen regions Figure 21 presents the arrangement of the screen areas discussed in the following Figure 4 6 Screen Areas The four screen areas offer the following feature...

Страница 37: ...The Info Health Status Area at the top of the display window appears in all screen displays Figure 4 7 shows an example of the Info Health Status Area on a unit that is in a proper working condition Figure 4 7 Info Health Status Area The Info Health Status Area displays the following information Unit Name top left in Figure 4 7 Date Time bottom left in Figure 4 7 as entered during the initial setu...

Страница 38: ...alth Status icon is not a touch sensitive area The yellow alert triangle is a Health Status Alert icon It indicates that there is an issue that does not stop operation of the biological safety cab inet immediately yet requires the operator s attention soon such as a HEPA filter to be replaced The Unit Health Status icon is a touch sensi tive area that opens a Health Status pop up window on top of ...

Страница 39: ...cal safety cabinet as can be seen in Figure 4 9 The Control Panel is visible at all times below the Main Screen 4 6 Model 1500 Series A2 Thermo Scientific Section 4 Operation Indicator Meaning The checkmark on the green circle indicates that this particular function is working properly The yellow animated Timer icon indicates that the unit is working up towards an operational state The white cross...

Страница 40: ...tch the device on including the blowers during the OFF Idle mode During the UV disinfection cycle the blowers run in reduced flow mode which is indicated by a small blue circle labeled 1 2 see lower image on the left side The fans also run in reduced flow mode when the front window is closed and there is no UV disinfection ongoing The Power Receptacle button switches the power receptacles in the s...

Страница 41: ...nifying that the airflows are not up to proper speed yet 4 8 Model 1500 Series A2 Thermo Scientific Section 4 Operation Icon Meaning Back button tapping this button will bring the user back up one level if one level or several levels down If there is no level to go back to this icon must be dark grey and must not be tappable If there is a level to go back to this icon must be blue and must be tapp...

Страница 42: ...d Figure 4 12 shows the unit in working mode with the window ready in the working position and the downflow and inflow speeds at operational speeds The sample space illumination and blowers are all on Figure 4 12 Main Screen Unit working Buttons are highlighted in the Control Panel Area to indicate that the sample chamber illumination and blowers are both on Power receptacle and UV functions are b...

Страница 43: ...cate that functions are still on including the blowers If the window is closed the blowers will switch to reduced flow as indicated by the small blue circle with the 1 2 label Idle mode is entered when you turn the fans off it is better for user to make sure that the window is either in the working position or closed Figure 4 14 shows that it is in Idle mode with the window closed Idle mode is ind...

Страница 44: ...The biological safety cabinet has a built in UV disinfection routine that affords an intensifying additional disinfection after a standard wipe spray disinfection UV disinfection can be started by pressing the UV button on the Control Panel or can be started automatically with the specified time if the UV auto start function is enabled For running the UV disinfection routine the front window must ...

Страница 45: ...isplayed on the Main Screen Figure 4 18 Main Screen UV disinfection running Additionally the Total cycle time in hours and the time remaining in hours minutes and seconds are displayed UV disinfection ends when the preset total cycle time for the routine has elapsed Then the UV lamps are switched off automatically and the UV disinfection screen goes away The UV disinfection can be started automati...

Страница 46: ...e Restart UV Cycle the disinfection will be restarted Figure 4 19 UV Cycle Interrupt when Window is open Tapping the Health Status Good icon in the Info Health Status Area takes you to the unit status shown in Figure 4 20 In this screen you can check the operating hours UV status and the Filter status Operating hours shows the total work hours UV shows the total UV work hours and the UV bulb insta...

Страница 47: ...the Reset button a pop up screen shown in Figure 4 21 Figure 4 21 UV Reset Confirm Pop up Screen Tap Yes Update button to save your setting you can also tap Cancel button to cancel your setting If you tapped the Yes Update button a pop up screen shown in Figure 4 22 Figure 4 22 UV Reset Saved Pop up Screen 4 14 Model 1500 Series A2 Thermo Scientific Section 4 Operation Replacing UV Bulb ...

Страница 48: ...with a Snooze button Additionally the Health status icon has a small blue circle overlay that shows the number of pending alarms NOTE Applicable BSC standards in effect may not allow temporary muting of audible alarm Tapping the Snooze button silences the audible alarm temporarily for a pre set timeout period The default alarm snooze timeout period is 5 minutes but the setting may be changed in th...

Страница 49: ...the problem proceed as follows Inspect the small blue alarm counter in the red Health Status Alarm icon to find out how many alarms you have to handle 4 16 Model 1500 Series A2 Thermo Scientific Section 4 Operation Information Displayed on Screen Alarm Type ER01 Inflow low Inflow alarm low ER02 Inflow high Inflow alarm high ER03 Downflow low Downflow alarm low ER04 Downflow high Downflow alarm hig...

Страница 50: ... has been cleared the screen display returns to the healthy state and the green Health Status Good icon replaces the red Health Status Alarm icon Some events in the biological safety cabinet that do not immediately affect containment or cleanliness produce alerts These alerts appear with the visual features shown in the example in Figure 4 26 Figure 4 26 Alert Screen The Health Status Alert icon i...

Страница 51: ...play the Status Alert screen shown in Figure 4 24 and review the alerts list Correct each issue that has caused an alarm As you correct the pending issues they will disappear from the alerts list on the Health Status Alert screen one by one When the last issue has been cleared the screen display returns to the healthy state and the green Health Status Good icon replaces the yellow Health Status Al...

Страница 52: ...you can set below 5 parameters Brightness The unit offers 4 brightness 25 50 75 and 100 50 brightness is the default setting Language The unit offers 2 languages for display English and Chinese English is the default setting Units of Measurement The unit offers 2 units for flow display metric units and imperial units metric units is the default setting metric units m s meters per second cmh cubic ...

Страница 53: ...w From the Alarms and Alerts screen you may change the audible alarms suspend time when Snooze button was tapped Minutes are selectable from a drop down list The default setting is 5 minutes You can tap the Save button to save the setting Figure 4 30 Alarms and Alerts Settings Screen 4 20 Model 1500 Series A2 Thermo Scientific Section 4 Operation Alarms and Alerts Snooze Time Setting Display Setti...

Страница 54: ...dditional intervals of 00 15 30 and 45 minutes Hours are entered by using a calculator style keypad minutes are selectable from a drop down list After UV cycle has been completed the default setting is to enter standby mode and you can also choose to enter idle mode by selecting the Idle mode after UV cycle has been completed NOTE The default setting is 1 hour 00 minutes 0 hours and 00 minutes is ...

Страница 55: ...latest records of alarm alert events Figure 4 32 Alarm Alert log Logfile Screen Tapping the Data Plate icon in the Settings screen takes you to the unit information shown in Figure 4 33 you can check the unit information model serial number main board firmware version GUI display board firmware version Figure 4 33 Data Plate Screen 4 22 1500 Series A2 Thermo Scientific Section 4 Operation Data Pla...

Страница 56: ...fety s 4 Load the work tray with samples 5 For work breaks or for extended experimental phases without manual intervention switch the device to standby mode by closing the window The observance of work rules ensures a minimum of operational safety when handling the biological safety cabinet Before starting a procedure take off all jewellry put on required personal protective equipment gloves goggl...

Страница 57: ...est should be 18 x 14 in 45 x 35 cm The slope should be adjustable to within a range from 5 to 15 The adjustable height should extend to a minimum of 4 in 11 cm above the floor After completing a procedure 1 Remove samples from the sample chamber and store them properly 2 Clean and disinfect the sample chamber surfaces including the work tray and the drain pan 3 Clean and disinfect all accessories...

Страница 58: ...n the front window is closed s The safety cabinet should remain connected to the power supply system at all times to ensure that settings for the individual unit configuration remain active in the memory If the power supply is interrupted and later the power supply connection has been reestablished the system switches to the operating mode that had been active last Additionally a window with the m...

Страница 59: ... device has been decontaminated close the window completely 2 Disconnect the unit from the power source All components with the exception of the HEPA filters can be discarded after having been thoroughly cleaned and decontaminated The HEPA filters must be discarded in accordance with the applicable national state and local regulations for special solid waste Warning As this unit can be used for pr...

Страница 60: ...sample chamber is required for the work procedure This sterilization procedure is mandatory before filters are replaced or before the unit is discarded Stainless steel is a combination of many different metals including iron and chromium Iron the primary element in stainless steel tends to corrode rust when in its natural state The chromium content in stainless steel prevents corrosion Stainless s...

Страница 61: ...neutralize the free iron deposits left on the steel where there has been deprivation of oxygen This stops the spread of the corrosion or rust It will not return the stainless steel to its original finish All the coated surfaces of this equipment are powder coated Powder coating is a method of applying a dry powder to electro statically charged metal then baked in an oven where the dry powder molec...

Страница 62: ... Be sure to dry completely 1 Rinse all surfaces twice with clean distilled water 2 Discard any liquid from the drain pan Rinse pan 3 Dry all chamber surfaces completely UV disinfection can be performed by using the optional factory installed UV light To start the UV disinfection procedure 1 Close the front window the air system operates in the reduced mode 2 Press the UV key on the control panel T...

Страница 63: ...ropriate Warning Decontamination with formaldehyde and chlorine dioxide must be performed in accordance with the specifications of NSF ANSI 49 Annex G As this procedure has considerable risks it must only be performed by specially trained and authorized service personnel Before Decontamination with chlorine dioxide the BSC must be sealed at the exhaust opening and front opening to eliminate exposu...

Страница 64: ...ompletely from the drain pan s 5 Re install the work tray The protective paper catch grids are retained in the airduct wall by their own tension 1 To remove a grid section push the retaining tab down until the section can be moved out from under the inner back wall 2 To install the grid section first place it on the lip of the back wall then push the retaining tab down and towards the back until t...

Страница 65: ......

Страница 66: ...LED bulbs are working Replace as necessary Replace UV bulb if necessary 1500 hour life Ensure the counterweight path is clear and unobstructed Ensure the sash is moving freely through its full range of travel Check sash cords and pullies for ware and or damage Inspect sash attachment points for cracks or damage Ensure all service valves are working properly Contact our service department if help i...

Страница 67: ...e rotate the bulb counterclockwise to disengage the latch and remove it from the sockets 4 To install slide the bulb contact pins into the rotating socket grooves and rotate the bulb clockwise to latch the sockets The factory installed UV light is installed in the chamber ceiling immediately behind the window The UV light should be replaced after 1500 operating hours See Figure 6 1 The UV light mo...

Страница 68: ...re cabinet with the exception of the HEPA filters can be discarded after having been thoroughly cleaned and decontaminated The HEPA filters must be discarded in accordance with the applicable national state and local regulations for special solid waste Warning As this unit can be used for processing and treating infectious substances it may become contaminated Prior to disposal the entire unit wit...

Страница 69: ......

Страница 70: ...ition mm in 254 10 pertains to all 4 6 models listed above with 10 window work position Opening max mm in 535 21 1 Working area height with adjustable height stand Height Range mm in 750 29 5 to 950 37 4 SmartPort Diameter mm in 76 0 3 0 Remaining side wall access ports Diameter mm in 23 0 91 SmartPort Distance lower edge mm in 365 8 14 4 Remaining upper ports Distance edge mm in 226 1 8 9 Remaini...

Страница 71: ...m temperature rise Above room temp with window closed K 2 2 Ergonomics Noise level dB A 4ft models 10 63 2 6ft models 10 65 2 The noise level was determined in accordance with NSF ANSI 49 The measurement uncertainty is within a range of 2 dB Volume weights and loads Description 4ft Models 10 window opening work position 1574 1584 6ft Models 10 window opening work position 1579 1589 Volume Drain pa...

Страница 72: ... current IEC 61010 1 UL61010 1 mA 3 5 Motherboard fusing A 2 x T 15 A Outlet fusing A 2 x T 5 A On site fusing A Circuit breaker 15 A Fuse T 15 A Power Power input maximum W 1130 1700 1440 2000 Protection Protection class I Protection type IP 20 Overvoltage category IEC 61010 1 UL61010 1 II Contamination degree IEC 61010 1 UL61010 1 2 Connecting lines Mains connection Power Cord 3 m 10 ft with NEM...

Страница 73: ... s 0 32 Downflow ft min 63 Air volume Overall volume flow m h 1415 2122 ft min 833 1249 Downflow volume flow m h 829 1244 ft min 488 732 Exhaust air volume flow m h 585 878 ft min 344 517 Filters Type HEPA Material Fiberglass Separability in MPPS 99 995 Separability at 0 3 μm particle size 99 999 Downflow Height mm in 457 18 Length mm in 1220 48 1830 72 Depth mm in 107 4 2 Exhaust air Height mm in...

Страница 74: ...n working with safety critical materials three to six months Nominal value Default value as specified by Thermo Measured value Value measured at the location of the safety cabinet Tolerance Acceptable deviation from the nominal value Average value The sum of the measuring values divided by the number of tests The average value is compared to the nominal value Setpoint Acceptable operating value fo...

Страница 75: ...st duct negative pressure and canopy performance As described in Annex F in addition to the above the following tests performed at the request of the customer or at the discretion of the certification provider Comfort and safety tests Lighting intensity Noise level Vibration Electrical leakage ground circuit resistance and polarity tests Note Service with costs Unless certification was expressly c...

Страница 76: ...fication Testing Test Equipment Testing equipment Manufacturer Application Smoke tubes MSA Pittsburgh Pennsylvania 15230 Airflow pattern test Thermoanemometer TSI Shoreview Minnesota 55126 Downflow velocity test and alternate method for inflow velocity using constricted access Digital Safety Inspector Ohmic Instruments Co Easton Maryland 21601 Electrical leakage and ground resistance tests GFI Cir...

Страница 77: ...V1 Inflow Air Volume V2 Actual Inflow Area A2 Acceptance 100 110 FPM Inflow Velocity Constricted Window Method Secondary Description Inflow velocity measured at several points through a constricted window Equipment Thermal Anemometer or equivalent hotwire Anemometer Probe Holder part number 1911325 8 4 1500 Series A2 Thermo Scientific Section 8 Certification Testing Testing Information continued ...

Страница 78: ... of the window glass 4 Beginning 3 inches in from the left inside wall record successive readings every 5 9 inches 8 readings for 4 ft and 12 readings for 6 ft models 5 Average those readings and correct result for true velocity 6 Calculate the inflow velocity as described below Calculation Inflow velocity V1 average of actual inflow velocity K factor 0 3 for 10 window opening work position K fact...

Страница 79: ...above the effective exhaust filter area in a grid pattern not to exceed 4 inches 10 cm x 4 inches 10 cm 4 Average the readings and calculate the inflow velocity Grid Pattern 4 ft Models 1584 1574 20 points 6ft Models 1589 1579 32 points Effective Exhaust Filter Area 4 ft Models 1584 1574 6ft Models 1589 1579 16 8 x 34 8 4 06 sq ft Inflow Area 4ft Models 10 window opening work position 1584 1574 Fr...

Страница 80: ...reestanding Fixture or equivalent Method 1 Operate the unit in the work position for approximately 20 minutes 2 Insert the anemometer probe into the probe stand adjusting for a height of 4 inches 10 cm above the bottom edge of the window opening while in the work position 3 Record the readings on a horizontal plane 4 inches 10 cm above the bottom edge of the window opening while in the work positi...

Страница 81: ...e the work tray and protective covers as appropriate 3 Introduce the aerosol from the generator in the center rear of the work area using a tee fitting to evenly distribute the aerosol 8 8 1500 Series A2 Thermo Scientific Section 8 Certification Testing Testing Information continued 6 5 5 5 5 6 6 5 88 5 88 5 88 5 88 5 88 5 88 5 88 5 88 5 88 5 88 6 inches 5 7 for 8 in opening models 33 points 6 5 5...

Страница 82: ...eter of the filter as prescribed in NSF ANSI 49 Annex F Acceptance Filters scanned 0 01 of upstream concentration at any point Method 1 Set up as in Steps 1 through 5 above 2 In the duct downstream of the filter sample in accordance with NSF ANSI 49 Acceptance Filters not scanned 0 005 of upstream concentration Description Determines the internal and external behavior of the air movement Equipment...

Страница 83: ...tion Verify the unit is integrated properly into the facility Description Verify the airflow alarms will activate within 15 seconds when there is a 20 loss Equipment Thermal Anemometer or equivalent hotwire Anemometer probe holder part 1911325 Shortridge ADM 870 or equivalent Flowhood series 8400 or equivalent Note To test the function of Airflow Alarms you need to access to the Field Certificatio...

Страница 84: ...w 1 inch 2 5 cm above work position Acceptance Window alarm activates Description Verify the exhaust system is functioning properly Canopy Connections Equipment Cold smoke titanium tetrachloride Method Pass smoke around the air gap to ensure negative pressure exists Acceptance No smoke refluxes back into the room once drawn into the canopy Sash Alarms Exhaust System Performance ...

Страница 85: ......

Страница 86: ...1500 Series A2 9 1 Thermo Scientific Device type Part number Serial number Service number Location Operator s note Technician Company Notes Date Signature Section 9 DataLog Warranty ...

Страница 87: ......

Страница 88: ...Thermo Fisher Scientific Suzhou Instruments Co Ltd No 297 Taishan Road New District Suzhou Jiangsu CN 215129 www thermofisher com ...

Отзывы: