Warning
Do not operate the unit before initial operation, installation and
certification tests have been performed.
s
The installation test must be performed in accordance with the
specifications of NSF/ANSI 49. The cabinet may be operated as a Class II
biological safety cabinet, in accordance with NSF/ANSI 49, if the unit
functions listed below were verified and if the test results are within the
safety value tolerances as specified in NSF/ANSI 49, Annex F.
• Inflow velocity profile test
• Downflow velocity profile test
• HEPA filter leakage test
• Airflow smoke pattern test
• Site installation assessment tests
• A repeat test must also be performed annually, after repairs to the unit
or after location changes.
• The operator should request a written test report from the authorized
service technician.
Warning
The operational safety of the unit, particularly the personnel and
product protection, are guaranteed only if all safety functions of the unit
have been tested and approved.
s
Caution
Thermo will not warrant operational safety if the unit is operated
without the required installation and certification test, or if these tests and
repeat test are not performed by adequately trained and authorized
personnel.
s
Caution
The initial operation with subsequent installation test does not
include any decontamination measures. The sample chamber and any
accessories required must be disinfected and cleaned in accordance with
the hygiene guidelines set forth for the desired application.
s
Caution
Alarms on this unit are not factory-set. Set the alarms to avoid
product loss and ensure personnel safety. Refer to Certification section.
s
1500 Series A2 3-5
Thermo Scientific
Section 3
Start-Up
Installation Tests